Chapter 2
General cells and cuvettes
Page 2-14
MAN 0487
Turn the cell upright and continue to inject slowly until the sample is reaches
the fill area as shown
. Fill between
shoulder of cell
and the
FILL MAX
line.
Check again for bubbles in the cell. Tap the cell gently to dislodge these.
Check that the electrodes are completely immersed.
ill 8796 / 8797
Remove the syringe and insert a cell stopper in
each
port.
Remove any liquid spilt on the electrodes contacts.
Note
The stoppers
must
be fitted before a measurement is performed. Ensure
that one stopper is fitted firmly, and the other one loosely, to avoid pressur-
isation of the cell.
DTS1060/61 cell
The cell name - DTS1060 - will be identified on the central section of the
cell body.
Prepare the sample in a syringe of at least 1ml capacity.
Place the sample syringe into one of the sample ports.
Invert the cell
.
Slowly
inject the sample from its syringe into the cell, filling the U tube to just
over half way
.
3