20
|
Auto contour
(not for Impulse Ambiente)
Use the corresponding buttons on the hand controller to set
auto contour.
(Auto contour operating buttons)
When this function is activated, the back and upper leg
sections are adjusted equally so that the occupant can be
brought into an upright sitting position.
This function should only be used with mobile occupants and
occupants without any physical problems.
Caution! Please consider occupant safety
when setting the auto contour. No foreign
objects should be located in the area of the
lifting mechanism.
Trendelenburg/
anti-Trendelenburg position
(not for Impulse XL)
The Impulse care bed range comes with reclining surface in-
clination as standard.
The head or foot height is set using the button on the hand
controller.
(Trendelenburg/anti-Trendelenburg
operating buttons)
Caution! Please consider the safety of the oc-
cupant when adjusting the bed inclination. No
foreign objects should be located in the area
of the lifting mechanism. This function must
only be activated by specialist staff and must
be locked on the back of the hand controller
by specialist staff. Incorrect settings may result
in serious injury to the occupant.
Hand controller locking function
The electrical unit combines state-of-the-art technology and
single-fault safety.
The locking function is a further safety precaution. The lock-
ing function is located on the back of the hand controller and
can be operated by staff with a key switch. In the event of
failure of the electrical motors, the functions can be disabled
via the corresponding rotary switch.
Operation:
The hand controller functions can be restricted by turning
the key switch to the different switch positions.
Symbols
F
Содержание Impulse Series
Страница 1: ...IMPULSE SERIES INSTRUCTIONS FOR USE CARE DESIGN...
Страница 2: ...2...
Страница 6: ...6 Impulse Ambiente Impulse XL Impulse 400 Impulse 400 LR 100 T H E P R O D U C T S...