47
MAINTENANCE
The manufacturer is only liable for the safety and reliability of the
product if it is serviced regularly and used in line with the safety
instructions. If any significant faults are found during maintenance
work which mean the safe operation of the product cannot be
guaranteed, the product must be taken out of use. Maintenance
work must be carried out at least once a year.
Any faults that impair the function and safety of the
hospital bed must be resolved before the bed is
used again and must be reported to the responsible
person.
Only original replacement parts from Betten Malsch
GmbH are permitted to be used.
Service and maintenance tasks must not be con-
ducted when the bed is occupied. The patient or
the care staff must not conduct maintenance tasks.
Procedure
1. Visual inspection
Check the welded structures for cracked weld seams and for
plastic deformation and wear. The welded structures include the
undercarriage and the sleeping surface with the moving interior
components. Also check that all screw joints are firmly connected.
2. Level of protection and functionality check of the side rails
During the functionality check, determine whether the side rails
can be easily locked in place and ensure that no impermissible
wear or deformation is visible.
The spacing requirements stated in BS EN 60601-2-52 are to
be used to check the level of protection. This is shown by the
following diagram and table.
Fig. 1 (dimensions of a single-piece side rail)
Содержание IMPULSE KL
Страница 1: ...CLINIC DESIGN IMPULSE KL INSTRUCTION MANUAL HOSPITAL BED...
Страница 2: ...2...
Страница 13: ...13...
Страница 31: ...31...
Страница 33: ...33 GS side rail dimensions When using the ABS metal lattice sleeping surface 378 114 85...
Страница 36: ...36 VGS side rail dimensions When using the ABS metal lattice sleeping surface 368 115 85...
Страница 41: ...41...
Страница 62: ...62 Notes...
Страница 63: ...63 Notes...