care bed and are based on the height adjustment and the
tilting motions of the bed. The minimum distance is 30 mm.
17. Improper use of the bed may cause hazards.
Examples of improper use are:
Unauthorised activation of the electrical functions
Use of the bed by children under the age of 12
Moving the bed by pulling on the mains cable
or side rails
More than one person adjusting the bed at
the same time
Activation of the functions by the occupant
without prior instruction
Pulling on the mains cable to disconnect it
from the power supply
Moving the bed on a sloping or rough surface
It is important that the following safety instructions are observed to
prevent risks to occupants and carers and to avoid any damage
to the bed:
In line with the standard 60601-2-52:2010, when
choosing a mattress it is important that there is a
minimum distance of 22 cm between the top of the
reclining surface and the top of the side rail in its fully
extended position. The mattress used must meet the
applicable safety standards.
Liquid permanently present in the motorised area must
be avoided (e.g. incontinence).
For safety reasons, the handle on the trapeze bar must
be replaced completely every five years.
Servicing and repairs on electrical components must
be carried out by specially trained staff and only origi-
nal replacement parts from the manufacturer must be
used.
The care bed is not suitable for extended operation
beyond a working cycle of two minutes. If the mains
adapter is overloaded or if it overheats, it will shut
off automatically. It can then only be operated again
after a cooling period of approx. 30 minutes. (Ob-
serve the drive manufacturer’s notes on the rating
plate!)
It is essential to avoid obstructing any part
of the bed mechanism as this can lead to damage
or complete disabling of the drive mechanism due to
overheating.
Likewise, the safe working load must not be exceeded.
If an immobile occupant remains in the same position
for a long period without the use of positioning aids,
this can lead to pressure sores. The manufacturer of
the care bed is not liable for this in any way.
Electrically operated care beds are medical products.
As such, they are subject to safety inspections in line
with the German Medical Device Directive (MPG) and