background image

8.  The thermal imager's 

FOV

 (Field of view) is 21 by 21 degrees.

9. 

FOV

 is the largest area that your imager can see at a set 

 

distance.

10.  This table lists the horizontal 

FOV

, vertical 

FOV

 and 

IFOV

 for 

 

the lense.

Focal Length

Horizontal FOV

Vertical

FOVIFOV

7.5mm

21°

21°

4.53mrad

IFOV

 (Instantaneous Field of View) is the smallest detail within the 

FOV

 

that can be detected or seen at a set distance, the unit is rad. The 
formula is this:

IFOV = ( Pixel Size )/ ( Lens focal length );

D:S (theoretical) [= 1/ 

IFOV

 (theoretical)] is the calculated spot size 

based on the pixel size of the Thermal Imager detector array and lens 
focal length.

Example: If the Thermal Imager uses a 9mm lens, the pixel size of the 
detector is 34um.

Horizontal 

FOV

 is 21°, Vertical 

FOV

 is 21°, the 

IFOV

 is 34um/7.5mm = 

4.53mrad;
D:S (theoretical) [= 1/ 

IFOV

 (theoretical)= 220:1]

MAX 10A

24

Содержание MT2005

Страница 1: ...MT2005 True RMS Thermal MultiMeter...

Страница 2: ...2...

Страница 3: ...7 4 13 AC DC Current Measurements 18 4 14 Using Range 19 4 15 Hold and AutoHold Mode 19 4 16 Capturing Minimum and Maximum Values 20 4 17 Related Values 20 4 18 Capturing Peak Values 21 4 19 Non Conta...

Страница 4: ...DC TRMS Voltage DC current AC AC DC TRMS current Resistance and Continuity test Diode test Capacity Frequency Duty Cycle Temperature with K type probe Flexible coils current Features 2 Safety This sym...

Страница 5: ...VOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient overvoltages to an appropriate low level OVERVOLTAGE CATEGORY II Note Examples include hou...

Страница 6: ...power supplies and disconnect the power when making resistance or diode tests ALWAYS turn off the power and disconnect the test leads before opening the covers to replace the fuse or batteries NEVER...

Страница 7: ...ary function switch 7 Positive Probe input jack for A Current 8 Positive Probe input jack for mA Current 9 COM Probe input jack 10 Positive Probe input jack for all Inputs except A and mA 11 Thermal m...

Страница 8: ...Navigation UP buttons to select PEAK function Use Navigation LEFT buttons to select REL function Use Navigation RIGHT buttons to select MAX function Cursor buttons Freezes the present reading in the d...

Страница 9: ...ystem s time 7 Indication of measuring unit 8 SD card 9 Temperature measuring result 10 Indication of Automatic Manual mode 11 Temperature unit 12 IR Camera 13 Indication of measuring unit 14 Indicati...

Страница 10: ...rement units and modifiers Button choices made in one function do not carry over into another function AC voltage measurements DC and AC DC voltage measurements Frequency and Duty measurements Resista...

Страница 11: ...itched ON or OFF Large voltage surges may occur that can damage the meter 4 2 AC DC Voltage Measurements 1 Set the function switch to the VDC position 2 Insert the black test lead banana plug into the...

Страница 12: ...e Frequency in the display 5 Press the MODE key to switch the Duty functions 6 Read the Duty in the display WARNING Risk of Electrocution The probe tips may not be long enough to contact the live part...

Страница 13: ...Insert the black test lead banana plug into the negative COM jack 3 Insert the red test lead banana plug into the positive Jack 4 Read the resistance in the display 4 5 Resistance Measurements To avoi...

Страница 14: ...ck 4 Press the MODE key to switch the Capacitance functions 5 Read the capacitance value in the Display 4 7 Diode Test 1 Set the function switch to the CAP position 2 Insert the black test lead banana...

Страница 15: ...ents 1 Set the function switch to the Flexible coil position 2 Insert the black test lead banana plug into the negative COM jack 3 Insert the red test lead banana plug into the positive V jack 4 Read...

Страница 16: ...nana plug into the A mA jack 3 For current measurements up to 600mA DC set the function switch to the mA position and insert the red test lead banana plug into the A mA jack 4 For current measurements...

Страница 17: ...set the function switch to the A position and insert the red test lead banana plug into the A mA jack 3 For current measurements up to 600mA AC set the function switch to the mA position and insert th...

Страница 18: ...set the function switch to the A position and insert the red test lead banana plug into the A mA jack 3 For current measurements up to 600mA AC DC set the function switch to the mA position and insert...

Страница 19: ...ot active in positions F Temp C F 10A 4 In Auto range mode the instrument selects the most appropriate ratio for carrying out measurement 5 If a reading is higher than the maximum measurable value the...

Страница 20: ...MAXIMUM readings unattended or recording readings while equipment operation precludes watching the Meter To activate the MIN MAX mode press the button If the Meter is already in MIN MAX function press...

Страница 21: ...ical circuits 1 The non contact voltage detector operates when the meter is set to any measuring function The detector does not operate when AUTO POWER OFF turns the meter off or when the rotary funct...

Страница 22: ...there is a SD card inserted 3 Bluetooth icon if this icon is displayed the Bluetooth is activated 4 The currently selected emissivity value use the Thermal Settings Menu to change the emissivity valu...

Страница 23: ...n the upper left hand corner for the targeted area along with the currently selected emissivity value 4 In the THERMAL IMAGING mode the laser pointer and display cross hairs can be used to assist in t...

Страница 24: ...the smallest detail within the FOV that can be detected or seen at a set distance the unit is rad The formula is this IFOV Pixel Size Lens focal length D S theoretical 1 IFOV theoretical is the calcul...

Страница 25: ...If the Meter is already in the MIN MAX function then Press the button to display the MIN value then press the button to display current measurement value Press once again to display the MAX value 3 Pr...

Страница 26: ...exit settings menu press MODE RANGE HOLD IR button or press the LEFT button in root menu 6 2 Settings details The Thermal imager has five kinds of palettes such as Press RIGHT MENU button to select on...

Страница 27: ...y detect the lowest temperature point 6 4 Measure Press the RIGHT MENU button to select this option While selected use the UP DOWN button to increase or decrease emissivity s value use the LEFT RIGHT...

Страница 28: ...er ON or OFF Brightness Press the RIGHT MENU button to select this option While selected use UP DOWN button to change LCD s brightness use the LEFT RIGHT MENU button to exit selection The available br...

Страница 29: ...the device For detailed information about Thermview please refer to Thermview APP help file Thermview for Android Search in Google Play with the keyword Thermview download and run Thermview for iOS Pl...

Страница 30: ...immediately Delete Photo 1 Press the RIGHT MENU button a dialog box will be displayed as show below Warning Selecting YES will delete all the photos in the memory card which captured by user Press th...

Страница 31: ...ss the LEFT RIGHT button to select previous or next picture Press any other keys to exit Image Browser mode 7 Image Browser 1 Left key instruction 2 Current displayed picture s filename 3 Right key in...

Страница 32: ...ent temperature 10 C 35 C object temperature 0 C Accuracy calculated as reading num digits resolution at 18 C 28 C 75 HR DC Voltage Range Resolution 600 0mV 0 1mV 6 000V 0 001V 60 00V 0 01V 600 OV 0 1...

Страница 33: ...olution 600 0uA 0 1uA 6000uA 1uA 60 00mA 0 01 mA 600 0mA 0 1mA Accuracy 50Hz 1 kHz Protection against overcharge 1 2 reading 5digits Quick fuse 800mA 1000V 10 00A 0 01A 1 5 reading 5digits Quick fuse...

Страница 34: ...RMS Protection against overcharge Accuracy 0 5 reading Sensitivity 2V RMS Frequency electronic circuits Range Resolution 60 00Hz 0 01 Hz 600 0Hz 0 1Hz 6 000kHz 0 001kHz 6 000MHz 0 001MHz Accuracy Pro...

Страница 35: ...1 2 reading 8digits 1 5 reading 8digits 2 5 reading 20digits Temperature with K type probe Range Resolution 40 0 C 600 0 C 0 1 C 600 C 1000 C 1 C 40 0 F 600 0 F 0 1 F Accuracy Protection against over...

Страница 36: ...l conditions for use Reference temperature Operating temperature Allowable relative humidity Storage temperature Storage humidity IEC EN61010 1 IEC EN 61326 1 Double insulation 2 CAT IV 600V CAT III 1...

Страница 37: ...V 3 0 3mV 3 0 3mV Accuracy Technical Specifications Note Accuracy is given as of reading counts of least significant digit at 23oC 5oC with relative humidity less than 80 RH Output Noise 5 5mV for eac...

Страница 38: ...ty Standard 1 5V AAA Size Battery X 2 Battery 2000m 7000ft maximum Operating Altitude 80 Storage Humidity Max 80 up to 31 C 87 F decreasing linearly to 50 at 40 C 104 F Operating Humidity 20 C to 60 C...

Страница 39: ...39...

Страница 40: ...illage 7 Sam Green Road Tunney Ext 9 Elandsfontein South Africa P O Box 888 Isando 1600 South Africa Telephone 27 11 872 5500 Sales Facsimile 27 11 822 2806 Admin Facsimile 27 11 822 1411 E mail sales...

Отзывы: