Automatic Measurement (AVM-510A)
9-18
SCHEDULED REPORT INTERVAL
Select this item and then turn the knob to set this selection for any time between 5 minutes and 480
minutes (8 hours). Reporting occurs at these intervals, starting immediately. The report ocurs whether
there is an error or not. If you set this item to 0 (OFF), just below the 5 minute point, periodic
reporting does not occur at all, and you will only see error reporting (if error reporting is enabled).
ERROR REPORTING
If selected, every time a new error occurs, another report is generated. This is in additon to any other
scheduled reports. When a new error occurs, the AVM-510A waits approximately 10 seconds before
sending a set of measurement data to a PC or printer. This prevents several, nearly simultaneous errors
from generating several printouts. However, the time of each error is noted next to it in the data set.
Note that each time a front panel control is moved or changed, the AVM-510A waits 20 seconds
before taking any measurements.
REPORT SUSTAINED ERRORS
When an error occurs, the AVM-510A normally generates a report. To keep generating reports while
the error is still there, set this option anywhere from 1 to 60 minutes. If this selection is set to 3
minutes, a new report is generated every 3 minutes, as long as the error keeps occuring. If this
selection is set to 0 (OFF), errors will be reported only when they first occur.
NOTE:
This function initiates a report after a set interval if ERROR REPORTING is
disabled.
REPORT ALL ERRORS CLEAR
If this selection is enabled, a special report is generated when all errors are cleared. All measurements
are given in this report. Note that if there are three errors and one of them clears, this condition is not
given in the special report.
NOTE:
This function operates only if ERROR REPORTING is enabled.
REPORTING DELAY
Lets you turn ON/OFF REPORTING DELAY. Refer to “ERROR REPORTING” described earlier in
this chapter.
About Reporting
1.
If the video signal is lost, a loss of video signal report is immediately generated and no
measurements are given. When video is restored, a reacquired video signal report is generated
along with the appropriate measurements.
2.
Out of tolerance conditions are registered immediately, but they are not reported to the printer or
PC for a few seconds. This enables the AVM-510A to make sure that all errors are reported. If the
AVM-510A does not wait, and 5 errors occurred within milliseconds of each other, a complete
report is generated for each of these errors, listing all the measurements each time.
3.
If any control on the front panel is changed, the AVM-510A waits 20 seconds before registering
an error. If the AVM-510A is currently reporting, do not change the front panel controls.
4.
Errors in a report are marked with an asterisk (*) and with the time they first occurred. Sustained
errors continue to carry this first occurance time in later reports. The time and date at the top of the
report shows when this particular report was generated.
Содержание AVM-510A Series
Страница 16: ...About This Manual 1 4 Notes...
Страница 34: ...Installation 4 6 Notes...
Страница 48: ...Front Panel Operation 5 14 Notes...
Страница 102: ...Viewing Waveforms Vectors and Pictures 7 30 Notes...
Страница 126: ...Automatic Measurement AVM 510A 9 20 Notes...
Страница 139: ...AVM 510A Option T For Transmission Monitoring 11 5 Figure 11 3 Differential Gain Figure 11 4 Differential Phase...
Страница 148: ...AVM 510A C For Component Measurements 12 4 Notes...
Страница 158: ...Connector Pinouts B 4 Notes...
Страница 162: ...Calibration D 2 Notes...