Automatic Measurement (AVM-510A)
9-17
In approximately 20 seconds you should see a light flashing on your printer signaling you that data is
being sent from the AVM-510A. You should start to see printed a series of measurements with the title
“New Error.” Every minute after this you should see printed a new list of measurements with the title
“Sustained Error.” If, after 5 minutes, you do not see anything from the printer, something is set up
wrong between the AVM-510A and the printer. Check the cabling and all the settings in the AVM-
510A menus and the printer menus.
Scheduling Reports
There are several options available to you when scheduling a report. The reporting process can operate
all the time or during a specified period of the day. Reporting can be at regular intervals, or when there
is an error, or both. Continuing errors can be reported at specified intervals. Afterwards, you can get a
special report after all errors have been cleared.
To schedule a report:
1.
Press the
Mesmnt
button.
2.
Select REPORT and then press the
Entr
button.
3.
The following REPORT menu appears on the display:
•
START/STOP TIME SETUP
•
SCHEDULED REPORT INTERVAL (0 off to 480 minutes)
•
ERROR REPORTING (ON/OFF)
•
REPORT SUSTAINED ERRORS (0 off to 60 minutes)
•
REPORT ALL ERRORS CLEAR (ON/OFF)
•
REPORTING DELAY (ON/OFF)
START/STOP TIME SETUP
Select an item in this menu to set it with the knob, press the
Entr
button, or drop into its sub-menus by
pressing the
Entr
button. The sub-menus found under this selection (by pressing the
Entr
button) let
you turn reporting on for a specific part of the day. If the START/STOP FUNCTION is off, the rest of
this sub-menu has no effect. If the START/STOP FUNCTION is on, set the start and stop time for
reporting by selecting the two lower slections (one at a time) and then by turning the knob for hours
and minutes. When START/STOP is on, the remainder of the REPORT menu is active only between
the start and stop times. When START/STOP is off, all remaining selections in the REPORT menu
work all the time.
START REPORT TIME
Select START REPORT TIME and then press the
Entr
button.
This selection lets you set (by turning the knob) a START REPORT TIME in hours and minutes by
selecting HOURS and then MINUTES.
STOP REPORT TIME
Select STOP REPORT TIME and then press the
Entr
button.
This selection lets you set (by turning the knob) a STOP REPORT TIME in hours and minutes by
selecting HOURS and then MINUTES.
Содержание AVM-510A Series
Страница 16: ...About This Manual 1 4 Notes...
Страница 34: ...Installation 4 6 Notes...
Страница 48: ...Front Panel Operation 5 14 Notes...
Страница 102: ...Viewing Waveforms Vectors and Pictures 7 30 Notes...
Страница 126: ...Automatic Measurement AVM 510A 9 20 Notes...
Страница 139: ...AVM 510A Option T For Transmission Monitoring 11 5 Figure 11 3 Differential Gain Figure 11 4 Differential Phase...
Страница 148: ...AVM 510A C For Component Measurements 12 4 Notes...
Страница 158: ...Connector Pinouts B 4 Notes...
Страница 162: ...Calibration D 2 Notes...