55
BE 58-602 Pulsar
®
Model R96 Radar Transmitter
Operational Mode
Current Consumption
Vmin
Vmax
HART
General Purpose
4mA
20mA
16.25V
11V
36V
36V
Intrinsically Safe
4mA
20mA
16.25V
11V
28.6V
28.6V
Explosion Proof
4mA
20mA
16.25V
11V
36V
36V
Fixed Current-Solar Power Operation (PV transmitter via HART)
General Purpose
10mA
➀
11V
36V
Intrinsically Safe
10mA
➀
11V
28.6V
HART Multi-Drop Mode (Fixed Current)
Standard
4mA
➀
16.25V
36V
Intrinsically Safe
4mA
➀
16.25V
28.6V
F
OUNDATION
Fieldbus
™
Supply Voltage
9V to 17.5V
9V to 17.5V
9V to 17.5V
3.7.2.2
Transmitter Terminal Voltage
3.7.3 O-ring (seal) Selection Chart
3.7.2.1
Safe Operating Area
0
16.25 V
24 V
36 V
11 V
➀
Start-up current 12 mA minimum
Material Code
Maximum
Temperature
Maximum
Pressure
Min.
Temp.
Recommended
For Use In
Not Recommended
For Use In
Viton
®
GFLT
0
+200 °C @ 16 bar
(+400 °F @ 232 psig)
51.7 bar @ +20 °C
(750 psig @ +70 °F)
-40 °C
(-40 °F)
General purpose, ethylene
Ketones (MEK, acetone),
skydrol fluids, amines, anhydrous
ammonia, low molecular weight
esters and ethers, hot hydrofluoric
or chlorosulfuric acids, sour HCs
EPDM
1
+120 °C @ 14 bar
(+250 °F @ 200 psig)
51.7 bar @ +20 °C
(750 psig @ +70 °F)
-50 °C
(-60 °F)
Acetone, MEK, skydrol fluids
anhydrous ammonia
Petroleum oils, di-ester base
lubricants, propane, steam
Kalrez
®
(4079)
2
+200 °C @ 16 bar
(+400 °F @ 232 psig)
51.7 bar @ +20 °C
(750 psig @ +70 °F)
-40 °C
(-40 °F)
Inorganic and organic acids
(including HF and nitric),
aldehydes, ethylene, glycols,
organic oils, silicone oils,
vinegar, sour HCs
Black liquor, hot water/steam, hot
aliphatic amines, ethylene oxide,
propylene oxide, molten sodium,
molten potassium
Simriz SZ485
(formerly
Aegis PF128)
8
+200 °C @ 16 bar
(+400 °F @ 232 psig)
51.7 bar @ +20 °C
(750 psig @ +70 °F)
-20 °C
(-4 °F)
Inorganic and organic acids
(including HF and nitric),
aldehydes, ethylene, glycols,
organic oils, silicone oils,
vinegar, sour HCs, steam,
amines, ethylene oxide,
propylene oxide
Black liquor, Freon 43, Freon 75,
Galden, KEL-F liquid, molten
sodium, molten potassium
Safe Operating Area
Digital Solar Mode
Vsupply
R
Loop
591
Ω
Typical HART
4-20 mA
Operating Area