REPLACEMENT PARTS

1

2

3

4

3

5

Item

Description

Part number

1

Electronic module

English display and HART

®

031-2809-001

French display and HART

®

031-2809-031

German display and HART

®

031-2809-041

Spanish display and HART

®

031-2809-021

HART

®

only

031-2809-002

English display only

031-2809-003

French display only

031-2809-033

German display only

031-2809-043

Spanish display only

031-2809-023

2

Terminal board

General Purpose (GP)

30-9106-003

Intrinsically Safe (IS)

30-9106-003

3

O-Ring (neoprene)

12-2201-237

4

Housing cover blind - Aluminium

04-9193-003

Housing cover blind - 316 SST

Consult factory

5

Housing cover with glass

36-4410-001

Housing cover with glass - 316 SST

Consult factory

8

APPROVALS

Agency

Description

ATEX

II 1G  EEx ia IIC T4 (ambient -40° C to +40° C)  
II 1G  EEx ia IIC T6 (ambient -40° C to +80° C)

FM (*)

Intrinsically safe 

Class I, Div.1; Grps. A, B, C, D

Class II, Div.1; Grps. E, F, G

Class III,NEMA 4X

Non-Incendive

Class I, Div.2; Grps. A,B,C,D

Class II, Div.2; Grps. F,G

Class III,NEMA 4X

(*) 

Consult factory for proper part n° and entity parameters.

Agency

Description

CSA (*)

Intrinsically safe 

Class I, Div.1; Grps. A, B, C, D

Class II, Div.1; Grp. G

Class III,Type 4X

Non-Incendive

Class I, Div.2; Grps A, B, C, D

Class II, Div.2; Grps E, F, G

Class III, Type 4X

(*) 

Consult factory for proper part n°

(*)

and entity parameters.

Содержание KOTRON SMART RF 805

Страница 1: ...R F Capacitance Level Measurement Installation and Operating Manual KOTRON SMART RF MODEL 805 TRANSMITTER...

Страница 2: ...and record the serial number for future reference when ordering parts MOUNTING Nameplate partnumber amplifier serial n tag n 2 cable entries one plugged M20 x 1 5 3 4 NPT or PG 13 5 Handtighten or us...

Страница 3: ...local current meter Ex Non Ex Ex Non Ex Ex Non Ex CAUTION ALWAYS check for proper grounding Improper grounding will cause malfunction of the unit CALIBRATION NOTE When connected to an approved barrie...

Страница 4: ...nt is the highest level from which the calibration is performed Hi Cal is measured from the end of the probe Lo Cal cm or inches Low Calibration point is the lowest level from which the cal ibration i...

Страница 5: ...single transmitter installation Fine tune the 4 mA point Attach a mA meter to the output If the output does not equal 4 0 mA adjust the value on the display to equal 4 00 mA Fine tune the 20 mA point...

Страница 6: ...oom Display Power Supply Current Meter 250 RL 450 CONNECTIONS Connection of your Hart communicator at power terminals and in wiring compartment at first junction box between unit and control room HART...

Страница 7: ...y HART software does not Contact manufacturer of your HART information e g Mfgs ID Device account for Mfgs ID s greater master software for upgrade This is a Serial etc but will not read any than 63 M...

Страница 8: ...3 003 Housing cover blind 316 SST Consult factory 5 Housing cover with glass 36 4410 001 Housing cover with glass 316 SST Consult factory 8 APPROVALS Agency Description ATEX II 1G EEx ia IIC T4 ambien...

Страница 9: ...Electronic Module Yes Yes Yes Yes Yes Yes Yes No No No No No No Display has Text No Reseat Boards Check Power Supply and Wiring Loop Resistance Wrong Correct Loop Resistance Start Good Display No Act...

Страница 10: ...lure to Operate with Correct Loop Inputs Display Working Yes No Retest Unit Good Display Key Board Responds Replace Electronic Module No Check Loop Current at Wiring Board Loop Current 22 4 0 mA Adjus...

Страница 11: ...heck Calibration Parameters Yes Yes Yes Corrupt Display Text Replace Electronic Module No Yes Replace Electronic Module Same Error Message Retest No Yes Active HART Analog Loop Correct Trim 4 20mA Che...

Страница 12: ...Material Aluminium A356T6 0 25 copper or 316 SST 1 4401 IP 65 Net and Gross Weight 2 70 kg net 3 20 kg gross Overall Dimensions H 214 mm 8 43 x W 111 mm 4 38 x D 188 mm 7 40 FUNCTIONAL PHYSICAL Descri...

Страница 13: ...e view Optional Mounting Flange 3 4 NPT 1 NPT 1G 1 BSP Optional Mounting Flange Optional Anchor Assembly Optional Weight 13 0 50 bare probe 16 0 625 insulated probe 5 0 1875 6 0 2500 75 2 95 34 1 32 1...

Страница 14: ...X II 1G EEx ia II C T6 A complete measuring system consists of a 1 Smart R F KOTRON 805 amplifier 2 KOTRON Probe a full range of rigid and flexible probes for conductive and non conductive media is av...

Страница 15: ...15...

Страница 16: ...Material Authorisation RMA form will be obtained from the factory It is mandatory that this form will be attached to each material returned This form is available through Magnetrol s local representat...

Отзывы: