TRANSMITTER SPECIFICATIONS

12

Description

Specification

Power (at terminals)

General Purpose: 11 to 36 V DC
ATEX Intrinsically Safe: 11 to 28.6 V DC

Signal Output

4-20 mA or 4-20 mA with HART

®

3.8 to 20.5 mA useable (meets NAMUR NE 43)

Span

minimum 5 pF - maximum 10.000 pF

Zero and Range

0 mm to 45 m (0' to 150') 

Resolution

Analog: 0.01 mA
Digital display: 0,1 cm (0.1")

Loop Resistance

ATEX Intrinsically Safe/General Purpose: 650 

@ 24 V DC

Damping

Adjustable 1-45 s

Diagnostic Alarm

Adjustable 3.6 mA, 22 mA, HOLD
(3.6 mA is not valid if unit includes HART

®

)

Digital Communication (HART

®

)

compatible with software version 5.x

User Interface

3-button keypad and/or HART communicator

Display

2-line x 8-character LCD

Approvals

ATEX II 1 G EEx ia IIC T4/T6 or FM/CSA intrinsically safe / non incendive

Electrical Data

Ui = 28,6 V, li = 140 mA, Pi = 1 W

Equivalent Data

Ci = 2,2 nF; Li = 0,037 mH

Menu Language

English/Spanish/French or German

Housing Material

Aluminium A356T6 (< 0.25% copper) or 316 SST (1.4401)/IP 65

Net and Gross Weight

2.70 kg net; 3.20 kg gross

Overall Dimensions

H 214 mm (8.43") x W 111 mm (4.38") x D 188 mm (7.40")

FUNCTIONAL/PHYSICAL

Description

Specification

Maximum Process Temperature

540 °C @ 35 bar (1000 °F @ 500 PSIG) - depending probe selection

Maximum Process Pressure

345 bar @ 40 °C (5000 PSIG @ 100 °F) - depending probe selection

Ambient temperature electronics

-40 °C to +80 °C (-40 °F to +175 °F) – blind transmitter
-20 °C to +70 °C (-5 °F to +160 °F) – with LCD display and keypad

Storage temperature

-40 °C to +80 °C (-40 °F to +175 °F)

Humidity

0-99 %, non-condensing

Electromagnetic Compatibility

Meets CE requirements (EN-50081-2, EN 50082-2) / NAMUR NE 21 (EMV)

PROCESS CONDITIONS

Description

Specification

Reference conditions

20 °C (70 °F)

Accuracy

+/- 0.5% of span or 2.5 mm whichever is greater

Resolution

0.1 pF

Repeatability

+/- 0.1% of span 

Linearity

+/- 0.25% of span

Response time

less than 1 s (adjustable via damping)

Ambient temperature effect

approxi0.03% per °C

Warm up time

less than 5 s

PERFORMANCE

1200

1000

800

600

400

200

0

20 mA

23 mA

24 V DC

620 

650 

11

0

10

20

30

40 V DC

OUTPUT TABLE

Содержание KOTRON SMART RF 805

Страница 1: ...R F Capacitance Level Measurement Installation and Operating Manual KOTRON SMART RF MODEL 805 TRANSMITTER...

Страница 2: ...and record the serial number for future reference when ordering parts MOUNTING Nameplate partnumber amplifier serial n tag n 2 cable entries one plugged M20 x 1 5 3 4 NPT or PG 13 5 Handtighten or us...

Страница 3: ...local current meter Ex Non Ex Ex Non Ex Ex Non Ex CAUTION ALWAYS check for proper grounding Improper grounding will cause malfunction of the unit CALIBRATION NOTE When connected to an approved barrie...

Страница 4: ...nt is the highest level from which the calibration is performed Hi Cal is measured from the end of the probe Lo Cal cm or inches Low Calibration point is the lowest level from which the cal ibration i...

Страница 5: ...single transmitter installation Fine tune the 4 mA point Attach a mA meter to the output If the output does not equal 4 0 mA adjust the value on the display to equal 4 00 mA Fine tune the 20 mA point...

Страница 6: ...oom Display Power Supply Current Meter 250 RL 450 CONNECTIONS Connection of your Hart communicator at power terminals and in wiring compartment at first junction box between unit and control room HART...

Страница 7: ...y HART software does not Contact manufacturer of your HART information e g Mfgs ID Device account for Mfgs ID s greater master software for upgrade This is a Serial etc but will not read any than 63 M...

Страница 8: ...3 003 Housing cover blind 316 SST Consult factory 5 Housing cover with glass 36 4410 001 Housing cover with glass 316 SST Consult factory 8 APPROVALS Agency Description ATEX II 1G EEx ia IIC T4 ambien...

Страница 9: ...Electronic Module Yes Yes Yes Yes Yes Yes Yes No No No No No No Display has Text No Reseat Boards Check Power Supply and Wiring Loop Resistance Wrong Correct Loop Resistance Start Good Display No Act...

Страница 10: ...lure to Operate with Correct Loop Inputs Display Working Yes No Retest Unit Good Display Key Board Responds Replace Electronic Module No Check Loop Current at Wiring Board Loop Current 22 4 0 mA Adjus...

Страница 11: ...heck Calibration Parameters Yes Yes Yes Corrupt Display Text Replace Electronic Module No Yes Replace Electronic Module Same Error Message Retest No Yes Active HART Analog Loop Correct Trim 4 20mA Che...

Страница 12: ...Material Aluminium A356T6 0 25 copper or 316 SST 1 4401 IP 65 Net and Gross Weight 2 70 kg net 3 20 kg gross Overall Dimensions H 214 mm 8 43 x W 111 mm 4 38 x D 188 mm 7 40 FUNCTIONAL PHYSICAL Descri...

Страница 13: ...e view Optional Mounting Flange 3 4 NPT 1 NPT 1G 1 BSP Optional Mounting Flange Optional Anchor Assembly Optional Weight 13 0 50 bare probe 16 0 625 insulated probe 5 0 1875 6 0 2500 75 2 95 34 1 32 1...

Страница 14: ...X II 1G EEx ia II C T6 A complete measuring system consists of a 1 Smart R F KOTRON 805 amplifier 2 KOTRON Probe a full range of rigid and flexible probes for conductive and non conductive media is av...

Страница 15: ...15...

Страница 16: ...Material Authorisation RMA form will be obtained from the factory It is mandatory that this form will be attached to each material returned This form is available through Magnetrol s local representat...

Отзывы: