3-2-1. Setting resolution
Th
e resolution denotes the diff erence in phase between the A and
B phase edges.
Th
e MJ100/110 can split scale signals from the detection head into
40 to 1000 divisions. Th
e default setting is 500 divisions (PL25:
10 μm; PL60: 4 μm). To change the resolution, set MODE switches
6 through 9 as given in Table 1.
Number of
Divisions
Resolution (µm)
MODE Switch
PL25
PL60
6
7
8
9
1000
5
2
ON
ON
ON
ON
960
OFF
ON
ON
ON
800
2.5
ON
OFF
ON
ON
512
OFF
OFF
ON
ON
500
10
4
ON
ON
OFF
ON
480
OFF
ON
OFF
ON
400
12.5
5
ON
OFF
OFF
ON
256
OFF
OFF
OFF
ON
240
ON
ON
ON
OFF
200
25
10
OFF
ON
ON
OFF
128
ON
OFF
ON
OFF
120
OFF
OFF
ON
OFF
100
50
20
ON
ON
OFF
OFF
80
OFF
ON
OFF
OFF
64
ON
OFF
OFF
OFF
40
125
50
OFF
OFF
OFF
OFF
Table 1
3-2-2. Setting the 1/2 division function
By using the 1/2 division function, it is possible to halve the
number of divisions (double the resolution) given in Table 1.
Th
e 1/2 division function setting is made using MODE switch 10.
MODE Switch 10
1/2 division function
ON
Do not halve the number of divisions
OFF
Halve the number of divisions
Th
e default setting is ON (do not halve the number of divisions).
Example: Resolution when the number of divisions is 100 at PL25
MODE Switch 10
Number of Divisions/Resolution
ON
100/50 µm
OFF
50/100 µm
Example: Resolution when the number of divisions is 500 at PL25
MODE Switch 10
Number of Divisions/Resolution
ON
500/10 µm
OFF
250/20 µm
Note
Th
e 1/2 division function (MODE switch 10) is not related to
the reference point synchronized with the A/B phase. For more
information, see the section “3-3. Using the reference point.”