![MagCore Plus II Скачать руководство пользователя страница 30](http://html2.mh-extra.com/html/magcore/plus-ii/plus-ii_service-manual_3402756030.webp)
━ Heater Test
━ UV Sterilization
Please Select Heater Test Button.
Please put the thermometer into the Heat Block
Well 1 and press Next.
Select Test Temperature
Now the heater will heat up to 65° C or 90° C,
please check whether the temperature shown on
the thermometer is consistent with what's shown
on the screen.
The acceptable temperature range is 65± 5° C or
90± 5° C.
Please Select the Sterilization Button.
Please select the sterilization time and press Start
button.
Warning!
The UV lamp is dangerous. Please do not
look straight at the inner of the machine
when it is operating.
Please stop operation before material
change.
Please do not open this cover by
untrained operators.
Please be cautious when UV light is on.
Содержание Plus II
Страница 1: ......
Страница 5: ...Specification Operating Parameters Operating Environment Applications...
Страница 6: ...IQ OQ PQDocuments IQDocuments...
Страница 7: ...OQ Documents...
Страница 8: ...PQ Documents...
Страница 9: ......
Страница 10: ...Accessories Installation 1 2 3 4 5 7 6 8 9 10 11 BeforeInstallation...
Страница 11: ...System Overview...
Страница 14: ...Installation...
Страница 15: ......
Страница 33: ...Replace Fuse Please use the specified type of fuse tominimizetheriskofburning...
Страница 35: ......
Страница 36: ...Disassemblethehousing 1 Loosen the two bottom screws to about 3mm...
Страница 37: ...2 Remove the remaining eight screws and take off the lid...
Страница 40: ...Axisdirections V axisdirection Y axisdirection M axisdirection V axisdirection X axisdirection...
Страница 42: ...ReplaceLEDlight 1 Remove the LED light from the top panel 2 Install a new LED light and hold it with cable ties...
Страница 49: ...Replace the sensor ofX axis 1 Remove the wire connected with the sensor 2 Remove the two screws on the sensor...
Страница 59: ...Replace Solid StateRelay SSR 1 Remove the two screws on SSR Remove the SSR and replace with a new one...
Страница 60: ...Electrical controlandengineeringsoftware...
Страница 61: ...Main board configuration...
Страница 66: ...10 Go to System Status and check the PLC version...
Страница 69: ...5 Go to System Status and check the HMI version...
Страница 72: ...6 Tap Confirmation when the program update is complete 7 Go to System Status and check the program version...
Страница 78: ......
Страница 79: ......