Troubleshooting
24
© Lutz-Jesco GmbH 2018
BA-40900-02-V01
Gas warning device
EASYCON GW
Operating instructions
11 Troubleshooting
See below for information about how to rectify faults on the device or the system. If you cannot eliminate the fault, please consult with the manufacturer
on further measures or return the device for repair.
Fault
Possible cause
Remedy
The device loses all settings after it has been
disconnected from the network and then
reconnected.
The battery is empty.
n
Check that the battery is really empty. To do
so, navigate to the menu System >
Information > System values. You can view
the battery voltage under “Battery”. If the
voltage is under 3 V, change the battery.
n
Replace the battery.
The device is off.
The power supply has been interrupted.
Restore the power supply.
The device fuse is defective.
Replace the fuse.
The sensor error is displayed as an alarm.
The sensor has not been installed correctly.
Ensure that the sensor has been installed
correctly. See section 7.5.
The sensor is not held in the sensor bracket
correctly.
Ensure that the sensor has been installed in the
sensor bracket correctly.
The signal cable to the sensor has a break.
Replace the sensor bracket.
Table 16: Troubleshooting