x
MAP Technology
Troubleshooting
39
There is a partial clog in the
system.
Clean the sample probe, then
flush the system. See page 20.
There is air in the system.
Verify sample probe height. Run
three Prime commands, two
Alcohol Flush commands, then
three Washes with water.
The waste container is not
vented.
Check that the seal in the waste
container cap is dry and that the
waste container cap is vented.
The waste line was moved
during system operation,
resulting in unstable flow rate.
Verify that the waste line is not
moved during system operation.
Possible problem with the laser.
View the calibration trend
report. Check for dramatic
changes in temperature, sheath
pressure, or voltages. If any of
these are evident on the report,
contact Technical Support.
Zero events collected during
calibration.
There is a problem with fluid
levels.
Check the sheath and waste
fluid levels. Verify that tubing
for both bottles is tightly
connected to the instrument.
Check that the waste bottle cap
is vented.
Laser-related issue.
Verify fluid is moving through
the system by performing a
Wash, which causes fluid to go
out to waste. If no fluid is going
to waste, clean the sample
probe, then flush the system
(see page 20). If the issue does
not resolve, contact Technical
Support.
The waste container is not
vented.
Check that the seal in the waste
container cap is dry and that the
waste container cap is vented.
The Cheminert fitting is loose.
Make sure that the Cheminert
fitting is tightened.
Analyzer fails Controls.
The control microspheres are
not fully suspended.
Vortex the control vials to
resuspend the microspheres.
Содержание Luminex 200
Страница 1: ...Luminex 200 System User Manual Lum nex...
Страница 9: ...iii Luminex 200 System User Manual Hardware 47 Software 48 xMAP Reagents 49...
Страница 10: ......
Страница 42: ......
Страница 56: ...Luminex 200 System User Manual 46...
Страница 60: ...Luminex 200 System User Manual 50...
Страница 64: ......