MAP Technology
Maintenance and Cleaning
19
9. Raise the sample probe using the Luminex software.
10. Replace the plastic shield that covers the sample probe area.
After Running Samples
After running samples:
1. Sanitize with a 10% to 20% household bleach solution.
2. Run two Wash cycles with distilled water.
3. Soak with distilled water. Wait until the soak completes.
4. If desired, turn off the Luminex 200 analyzer.
Routine Tasks
Sheath and Waste Fluids
Replace the sheath fluid and empty the waste container as required.
Be careful not to touch the seal in the waste bottle cap or allow it to
become wet or dirty. This could cause the bottle to become
pressurized, causing pressure errors in the system. If the seal gets
wet, let it air-dry. Touching the seal can contaminate it. If it gets
dirty, you should replace the waste bottle cap.
If you are using the waste line, do not move it while the system is
running. While you can move the line along a horizontal surface, do
not permanently reroute the waste line elevation without first
contacting Technical Support. You can move the line temporarily for
cleaning and maintenance purposes.
You must manually monitor waste container levels.
Refill the Sheath
Fluid Container
To refill the sheath fluid container:
1. Release system pressure by removing the lid from the sheath
fluid container.
2. Refill the sheath fluid container.
If the sheath container ever runs dry, prime the system at least twice
until the air is removed from the system.
Содержание Luminex 200
Страница 1: ...Luminex 200 System User Manual Lum nex...
Страница 9: ...iii Luminex 200 System User Manual Hardware 47 Software 48 xMAP Reagents 49...
Страница 10: ......
Страница 42: ......
Страница 56: ...Luminex 200 System User Manual 46...
Страница 60: ...Luminex 200 System User Manual 50...
Страница 64: ......