background image

IMPAC pyrometers IN 300 · IN 300/5

 

 

 

 

18 

5

 

Optics 

 
The pyrometers are equipped ex works with one of the following optics. These lenses are focusing 
to certain distances, i.e. in these distances each lens achieves its smallest spot size in relation to 
the measuring distance. The spot size will change in any other distance (shorter or longer). Please 
notice that the measuring object must be as least as big as the spot size.  
The name of the optics (e.g. optics 300) shows the measuring distance in mm (e.g. 300 mm) in 
which it has the smallest spot size in relation to its measuring distance (IN 300, e.g. 20 mm). 
The following drawings show the size of the spots in mm in dependence of the measuring dis-
tance. Values between the mentioned data can be calculated by interpolation. The spot size for 
measuring distance 0 is the aperture diameter of the optics.  
 
 

5.1 

IN 300 

 
 
 

Optics 300: 

 

Spot size [mm]

Measuring distance [mm]

0

8

150

300

600

1000 

85 

48 

20 

14 

 

 
 

5.2 

IN 300/5 

 
 

 

Optics 80: 

 

5

0

80 

200 

14 

2,5

Spot size [mm]

Measuring distance [mm]

 

 
 
 

 

 

Optics 160: 

 
 

 

5

0

160

230 

10 

Spot size [mm]

Measuring distance [mm]

 

 
 

(lens tolerances are not taken into consideration) 

 

Содержание IN 300

Страница 1: ...IN 300 IN 300 5 IMPAC Pyrometer Betriebsanleitung Operation Manual Proven Quality www imaxsun com...

Страница 2: ...ons 15 2 2 Appropriate use 16 2 3 Scope of delivery 16 3 Safety 16 3 1 General 16 3 2 Electrical connection 16 4 Electrical Installation 16 4 1 Accessories 17 5 Optics 18 5 1 IN 300 18 5 2 IN 300 5 18...

Страница 3: ...nd experience LumaSense Technologies is not liable for any damages that arise from the use of any examples or processes mentioned in this manual or in case the content of this document should be incom...

Страница 4: ...tput 4 20 mA load independent current linear to temperature Max load max 500 at 24 V max 200 at 18 V Emissivity 0 4 1 adjustable Repeatability 1 of the measuring range 1 Tamb 23 C Accuracy 1 5 of temp...

Страница 5: ...connection Follow common safety regulations for mains voltage 230 or 115 V AC connecting additional devices operating with this mains voltage e g transformers Touching mains voltage can be mortal A no...

Страница 6: ...C Recorder Digital display 230 V 4 20 mA white brown Shield black 24 V DC Power supply Wiring white 24 V brown 0 V black Shield Note Additional analyzing instruments e g controllers recorders etc can...

Страница 7: ...the measuring distance in mm e g 300 mm in which it has the smallest spot size in relation to its measuring distance IN 300 e g 20 mm The following drawings show the size of the spots in mm in depende...

Страница 8: ...is set ex works to an emissivity of 100 To adjust the emissivity factor to your own purpose you have to remove the rear cover by unscrew the locking screw Remove the cover carefully In the tube there...

Страница 9: ...unction the pyrometer 8 Maintenance 8 1 Safety Attention during pyrometer services Should the pyrometer be integrated in a running machine process the machine has to be switched off and secured agains...

Страница 10: ...8 2 Temperature indication too high Emissivity set too low Set lower correct emissivity corresponding to the material see 6 The measurement is influenced by reflections of hot machine parts Use mecha...

Страница 11: ...DA 4000 LED display 2 wire power supply 2 limit switches relay contacts 3 890 520 DA 6000 LED display RS232 2 wire power supply maximum value storage analog output 3 890 530 DA 6000 LED display RS485...