Fiber optic
Depending on the measuring range
two different fibers are used. They are
marked red or blue. The color mark
must be mounted on the pyrometer’s
side.
The fiber optic is equipped with a
special connector which allows to fix
the fiber only in one position. Please
notice that the connector of the fiber
with the pin is fixed in the correct
position before plug and socket can be
bold together.
3.4.3
Color mark at the fiber
Blue: MB 13
Red: MB 18, 25
3.4.4
Ambient temperature
Fiber and optical head can withstand ambient temperatures up to 250 °C without cooling on
optical head’s side.
3.4.5
Serial number
The original fiber has a serial number which is also on the pyrometer’s housing. If required, the
fiber can be replaced by a new original IMPAC fiber. A recalibration of the pyrometer is not
necessary if it is only used in ratio mode or pouring stream mode.
If the pyrometer is also used in
mono mode or metal mode it must be calibrated together with the fiber (service).
3.4.6
Minimum bending radius
Blue: 100 mm for short time, 300 mm permanently
Red: 50 mm for short time, 120 mm permanently
Note:
A hot fiber optic cable should not be exposed to continual movement!
3.4.7
Optical head
The FSMA connector on the side of the optical head
doesn’t need a special justification if bold together.
If required, the optical head can be replaced by
another one (also without line spot size). A
recalibration of the pyrometer is not necessary if it is
only used in ratio mode.
If the pyrometer also used in
mono mode or metal mode it must be calibrated
together with the fiber (service).
Fiber
Screw connection
Pin
Cut-out
Color mark