background image

 

ISR 12-LO/GS Manual 

Controls and Installation 

  16 

 
 

 
 

3.1.2

 

Connector pin J 

The connector pin J can be used for 4 different functions: 

1) Operating mode contact: 

During the warm-up of the pyrometer (after connection to the 

power supply, the LED display on the converter indicates  “7777”) pin J is connected to the 
power supply voltage. This voltage is connected by a relays switch (max. continuous current  
0.4 A) to pin K (power supply voltage). This relays switch opens when the warm-up is finished, 

and the pyrometer is in operating  mode. For that reason,  this pin can be used as operating 
mode contact. 

 

2) External clearing of the maximum value storage:

 When the pyrometer is in operating 

mode, pin J can be used for external clearing of maximum value storage (see

 4.7

). To clear the 

maximum value storage, connect pin J for a short time to pin K (power supply voltage). 
The function “external clearing” is activated with the following conditions: 

 

Warm-up is finished (no display “7777“) 

 

The clear time is set to “extern“ (see 4.7) 

 

The lens contamination monitoring is switched off. This can be done via software InfraWin 

in “color-bar“ window. The warning level “dirty window” must be set to 0%. 

3) Lens contamination monitoring system:

  The pyrometers are equipped with a 

contamination monitoring system. Contamination of lens or sighting window or dust in the 
sighting path of the pyrometer can interfere the signal in such a high level that a correct 
temperature measurement will be impossible (this can also happen if the object is smaller than 
the spot size of the pyrometer). To avoid wrong measurements in advance, a warning signal can 

be set to a certain contamination level. If contamination reaches this level (or at a certain too 
low signal level), a built-in relays (max. continuous current 0.4 A) connects pin J to pin K (power 
supply voltage). The setting of the switch level (0 … 99%) can be done via software InfraWin in 
“color-bar“ window. If the warning level “dirty window” is set to 0% the lens contamination 

monitoring system is switched off and pin J has the function “external clearing” of maximum 
value storage (factory setting). 
The lens contamination monitoring system is activated with the following conditions: 

 

Warm-up is finished (no display “7777“) 

 

The clear time is not set to “extern“ or “hold” (see 

4.7

). 

 

The pyrometer is operating in “2-color mode” (see

 4.8

4) Extern Hold: 

When the hold function mode is activated the current pouring stream 

temperature 

(or the current temperature reading if the instrument is not in pouring stream 

mode) 

is frozen if J and pin K are connected (see 

4.7 clear time of the maximum value 

storage

). 

3.2

 

Connecting the pyrometer to a PC 

The pyrometers are equipped with a serial interface RS232 or RS485 (switchable at the 
pyrometer). Standard on a PC is the RS232 interface. At this interface one pyrometer can be 
connected if the interface is set to RS232. Only short distances can be transmitted with RS232 

and electromagnetic interferences can affect the transmission.  
 
With RS485 the transmission is to a large extend free of problems, long transmission distances 
can be realized,  and several pyrometers can be connected in a bus system. If RS485 is not 

available at the PC, it can be realized with an external converter which converts the RS485 in 
RS232 for a standard connection to a PC.  

Содержание IMPAC ISR 12-LO/GS

Страница 1: ...MANUAL IMPAC Pyrometer ISR 12 LO GS...

Страница 2: ...nse Technologies All other trademarks are trademarks registered trademarks and or service marks of their respective holders Service Centers LumaSense Technologies Inc North America Sales Service Santa...

Страница 3: ...ing 13 3 Controls and Installation 15 3 1 Electrical Installation 15 3 1 1 Pin assignment for the male socket on the bottom side of the pyrometer 15 3 1 2 Connector pin J 16 3 2 Connecting the pyromet...

Страница 4: ...perating mode 1 2 M 29 4 9 Switch off limit OFF 30 4 10 Analog Output 30 4 11 Sub Range 30 4 12 Device Address 30 4 13 Baud Rate 31 4 14 Analog Test 10 mA 31 4 15 Wait time tw 31 5 Software InfraWin 3...

Страница 5: ...ty note laser beam Indicates to the danger of a built in laser targeting light Text Text written in gray color explains integrated functions of the pyrometer which are not used in pouring stream mode...

Страница 6: ...ondition exists that would cause the software not to run smoothly The manufacturer assumes no responsibility or liability and will not guarantee the performance of the software Liability regarding any...

Страница 7: ...eight damage to instruments that are improperly packed Contact us to obtain an RMA number if Technical Support has not already assigned one Clearly indicate the assigned RMA number on the shipping pac...

Страница 8: ...ISR 12 LO GS Manual General Information 8 1 8 Disposal decommissioning Inoperable instruments must be disposed of in compliance with local regulations for electronic materials...

Страница 9: ...antees that even a moving pouring stream is always within the spot The pouring stream temperature of each single casting process and with that of each single casted part can be recorded automatically...

Страница 10: ...ues in which must exceed the beginning of the adjusted temperature range Environment Protection Class IP 65 DIN 40 050 Ambient Temperature Converter housing 0 60 C Optical head up to 250 C Storage Tem...

Страница 11: ...s 25 s extern via interface or automatically with the next measuring object hold function to freeze the actual displayed pouring stream temperature Electrical Power Supply 24 V DC 15 40 V DC or 24 V A...

Страница 12: ...ISR 12 LO GS Manual Introduction 12 2 4 Dimensions Pyrometer housing Optical head cover window Lens all dimensions in mm...

Страница 13: ...cal fiber 15 Mounting holes 2 6 Accessories Optional Numerous accessories guarantee easy installation of the pyrometers The following overview shows a selection of suitable accessories You can find th...

Страница 14: ...ts the lens from contamination with dust and moisture It must be supplied with dry and oil free pressurized air 1 5 m h and generates an air stream shaped like a cone Displays Additionally to the buil...

Страница 15: ...cting cable has wires for power supply interface analog output external laser switch and external clear of maximum value storage via contact see section 8 Reference numbers and 12 pin connectors The c...

Страница 16: ...meter To avoid wrong measurements in advance a warning signal can be set to a certain contamination level If contamination reaches this level or at a certain too low signal level a built in relays max...

Страница 17: ...Bd is 7 m PC s side soldering side K A B C D J G F E H M L 1 2 3 4 5 6 7 8 9 DGND red RxD black TxD violet Pyrometer s side soldering side 3 2 2 Connecting to RS485 interface Half duplex mode A1 and A...

Страница 18: ...f 1 mW The laser is classified as product of laser class II Warning To reduce the risk of injury to the eyes do not look directly into the targeting laser and do not point the targeting laser into any...

Страница 19: ...housing If required the fiber can be replaced by a new original IMPAC fiber A recalibration of the pyrometer is not necessary if it is only used in ratio mode or pouring stream mode If the pyrometer i...

Страница 20: ...above shows some examples for different measuring distances distance measured from front of the lens to the object see 2 4 Dimensions 3 5 2 Spot size table Optical head with line spot width Optical h...

Страница 21: ...size of this line varies for the different measuring distances As the alignment of the optical head is quite difficult during the pouring process it is recommended to do it before the process starts...

Страница 22: ...ISR 12 LO GS Manual Controls and Installation 22 To ensure consistent document formatting this page was intentionally left blank...

Страница 23: ...g of other modes only via software InfraWin see Chapter 5 Software InfraWin 4 1 Factory Settings Pouring stream mode On Start up condition 50 Pre run time 0 5 s Measuring time 0 auto Follow up time 0...

Страница 24: ...activates a test current of 10 mA on the analog output independently if the instrument is adjusted to 0 to 20 mA or 4 to 20 mA This enables to test external indicators for correct scaling The readings...

Страница 25: ...ring stream temperature is available Display blinking Signal strength in foundry mode is below the adjusted level 4 3 Pouring stream mode The start and end processes of the pouring disturb the tempera...

Страница 26: ...process automatically and is using this period for the calculation of the temperature and shows the calculated temp on the display In both modes the temperature is displayed until the next calculated...

Страница 27: ...ature by ratioing the radiation intensities of the two wavelengths This ratio technique eliminates a number of factors that degrade the accuracy of a conventional single color instrument e g measureme...

Страница 28: ...tered time the storage will be deleted Extern The external clearing can be activated and used within an own software see section 7 Data format UPP or via an external contact for connection see 3 1 1 P...

Страница 29: ...age stores always the highest value during the measurement value calculation Setting a value for the exposure time is smoothing this stored value so that even with interruptions in the pouring stream...

Страница 30: ...ing range of the pyrometer This subrange corresponds to the analog output from describes the beginning of this measuring range to the end of the range For automatic detection of a pouring stream the b...

Страница 31: ...ading of the internal indicator should correspond to the reading of the external indicator If both displays show different readings the external indicator has a wrong temperature range or input curren...

Страница 32: ...ISR 12 LO GS Manual Settings Parameter Descriptions 32 To ensure consistent document formatting this page was intentionally left blank...

Страница 33: ...com 5 1 Connecting the pyrometer to a PC The program InfraWin can operate up to two devices Two devices using RS485 may be operated simultaneously by the same interface if two different addresses have...

Страница 34: ...ISR 12 LO GS Manual Software InfraWin 34 To ensure consistent document formatting this page was intentionally left blank...

Страница 35: ...only the lens must be kept clean The lens can be cleaned with a soft cloth in combination with alcohol do not use acid solutions or dilution Also standard cloths for cleaning glasses or photo objectiv...

Страница 36: ...ISR 12 LO GS Manual Maintenance 36 To ensure consistent document formatting this page was intentionally left blank...

Страница 37: ...7 or 97 0 The ASCII parameter X indicates a change to be made in a parameter When used the command contains the new value Example Write Command Entry 00emXXXX CR The parameter used for the emissivity...

Страница 38: ...hex 8 digit C or F XXXX beginning of temperature range YYYY end of temperature range Reading temperature sub range AAme Same as mb Setting of temperature sub range AAm1XXXXYYYY XXXX hex 4 digit begin...

Страница 39: ...number AAsn Output XXXX hex 4 digit Device type software version AAve Output XXYYZZ 6 digit decimal XX 06 ISR 12 LO YY month of software version ZZ year of software version Software version in detail...

Страница 40: ...ISR 12 LO GS Manual Data format UPP Universal Pyrometer Protocol 40 To ensure consistent document formatting this page was intentionally left blank...

Страница 41: ...t switches 3 852 440 Protocol transducer RS485 RS232 switch Profbus DP for 1 device 3 852 460 Protocol transducer RS485 Profbus DP for 32 devices 3 852 620 Protocol converter UPP RS485 or RS232 ProfNe...

Страница 42: ...ISR 12 LO GS Manual Reference Numbers 42 To ensure consistent document formatting this page was intentionally left blank...

Страница 43: ...s Use mechanical construction to avoid the influence of the interfering radiation sighting tube No temperature indication at the end of the pouring stream Pouring stream is not in the spot of the pyro...

Страница 44: ...ISR 12 LO GS Manual Troubleshooting 44 To ensure consistent document formatting this page was intentionally left blank...

Страница 45: ...mation 5 I Installation Electrical 15 L Language 33 Laser targeting light 18 Legend 5 Lens contamination monitoring system 16 Liability 6 M Maintenance 35 Maximum value storage 16 Measuring condition...

Страница 46: ...ot Sizes 20 Storage Modes 28 Sub Range 30 Support 6 Switch off limit 30 T Technical Data 10 Temperature display 24 27 Temperature sub range 25 Test current 24 Troubleshooting 43 U Unpacking the Instru...

Отзывы: