ISR 12-LO/GS Manual
Settings / Parameter Descriptions
•
24
4.2
Instrument settings
Display °C or °F
Parameter indicator:
Emissivity
/
emissivity slope (
ε
/ K)
Exposition time (t
90
)
Clear time of maximum value storage (t
CL
)
Operating mode(1/2/M)
Switch-off limit (OFF)
Current output 0 or 4 ... 20 mA (0/4 mA)
Beginning of sub range (from)
End of sub range(to)
Address (ADR)
Baud rate (kBaud)
Temperature display in °C or °F (°C/°F)
Push-button for test current
LEDs for clear mode
Measuring mode display:
-
mono mode
-
ratio mode
-
metal mode
Temperature- or
parameter display
Setting keys
Targeting light on / off
Interface
RS232 or RS485
Scale °C or °F:
The LED indicates the temperature scale in °C or °F.
Parameter indicator:
LED’s indicate which pyrometer parameter is selected for reading or
changing.
Push button for test
current:
The test button activates a test current of 10 mA on the analog output,
independently if the instrument is adjusted to 0 to 20 mA or 4 to 20 mA.
This enables to test external indicators for correct scaling. The readings of
the internal indicator should correspond to the reading of the external
indicator. If both displays show different readings the external indicator
has a wrong temperature range or input current. This test current can be
switched off by pushing the test button again or any other setting key or
it is switched off automatically after 30 seconds and the instrument is
working in the measuring mode.
LED for
clear mode:
LED indicates if the maximum value storage is operating in automatic or
external clearing mode.
Temperature or
parameter display:
In the pouring stream mode, the display always shows the last pouring
stream temperature,
in other modes the display shows the actual
temperature reading
. After pushing the parameter key the display
indicates the actual value of one selected parameter.