ISR 320 Manual
Data format UPP
•
29
7
Data Format UPP (Universal Pyrometer
Protocol)
Software commands can be exchanged directly with the pyrometer through an interface using
suitable communication software or by using the
Test
function located in the
Pyrometer
Parameters
window of the InfraWin software package.
The data exchange occurs in ASCII format with the following transmission parameters:
•
The data format is: 8 data bits, 1 stop bit, even parity (8,1,e) no handshake;
•
The device responds to the entry of a command with output (such as the measuring
value) + CR (
C
arriage
R
eturn, ASCII 13), to pure entry commands with
ok
+
CR
, or
no
+
CR
.
•
Every command starts with the 2-digit device address AA followed by two lower case
command letters and finished with CR.
o
Example Read Command:
Entry: “00em“ + CR
o
The emissivity setting (
ε
) of the device with the address 00 is returned.
o
Answer: “0970“ +
<
CR
>
means Emissivity = 0.97 or 97.0%
•
The ASCII parameter “X” indicates a change to be made in a parameter. When used, the
command contains the new value.
o
Example Write Command:
Entry: “00emXXXX“ + CR
o
The parameter used for the emissivity setting (
ε
) with the address 00 is changed.
o
Answer: “00em0853“ +
<
CR
>
changes the Emissivity to 0.853 or 85.3%
•
A “?” after the lower case command letters answers with the limits of the respective
settings (only at setting commands, not at query commands).
Example Read Limits Command:
Entry: “00em?“ + ˜CR!
Answer: Could be 02 <CR>, which means that E can vary between 0.200 and
1.000 (or 20% and 100%)
Description
Command
Parameters
Analog Output
AAasX
X=0 (0 to 20 mA)
X=1 (4 to 20 mA)
Minimum Intensity
Switch-Off Level
AAawXX
AAaw
XX=02 to 50
(2% to 50%)
Answer: DD
2 decimal digit 02 to 50
Reference Number
AAbn
Output: XXXXXX (hex 6-digit)
Baud Rate (set)
AAbrX
X = 0 to 6 or 8 (dec.)
0 = 1200 Baud 1 = 2400 Baud 2 = 4800 Baud
3 = 9600 Baud 4 = 19200 Baud 5 = 38400 Baud
6 = 57600 Baud 7 = is not allowed 8 = 115200 Baud
“Dirty Window”
Warning
AAdwXX
XX = 00 …99% (2 digit, hex.)
Содержание ISR 320
Страница 1: ...MANUAL IMPAC Pyrometer ISR 320 ...