ISR 12-LO
•
IGAR 12-LO Operating Manual Data format UPP (Universal Pyrometer Protocol)
•
36
Description
Command
Parameters
Clear times of
maximum value
storage:
AAlzX
X = 0 ... 8 (dec.) 0 = Maximum value storage off
1 = 0.01 s
4 = 1.00 s
7 = external deletion
2 = 0.05 s
5 = 5.00 s
8 = automatically
deletion
3 = 0.25 s
6 = 25.00 s
9 = hold
External deletion:
AAlx
Simulation of an external deletion contact
Operation mode:
AAkaX
X = 0 metal mode from channel 1 and 2
X = 1 mono mode
X = 2 ratio temperature
Switch-off limit:
AAawXX
XX = 02 ... 50%
Analog output:
AAasX
X = 0 or 1
0 = 0 to 20 mA 1 = 4 to 20 mA
Reading basic
temperature range:
AAmb
Output: XXXXYYYY (hex 8-digit, °C or °F)
XXXX = beginning of temperature range
YYYY = end of temperature range
Reading temperature
sub range:
AAme
same as mb
Setting of
temperature sub
range:
AAm1XXXXYYYY XXXX (hex. 4-digit) beginning of temp. range
(°C or °F)
YYYY (hex. 4-digit) end of temp. range (°C or °F)
Address:
AAgaXX
XX = (00 ... 97)
00 ... 97 = regular device addresses
99 = global address with response
98 = global address without response (only setting
commands!)
Baud rate:
AAbrX
X = 0...6 or 8 (dec.)
1 = 2400 baud 4 = 19200 baud (7 is not allowed)
2 = 4800 baud 5 = 38400 baud 8 = 115200 baud
3 = 9600 baud 6 = 57600 baud
Changing °C / °F:
AAfhX
Output: X = 0: display in °C; X = 1: display in °F
Reading signal
strength:
AAtr
Output: XXXX, XXXX = 0000 ... 9999 in ‰
Warning level:
AAdwXX
XX = 00 ... 99% (2 digit, hex.)
Wait time:
AAtwXX
XX = 00 ... 99 (decimal)
Error status:
AAfs
Output: XX;
XX=00 … FF (00 = no error)
(01 … FF: error code for LumaSense service)
Internal temperature: AAgt
Output: XXX (dec. 000 ... 099 °C or 032 ... 210 °F)
Max. internal
temperature:
AAtm
Output: XXX (dec. 000 ... 099 °C or 032 ... 210 °F)
Reading interface
type:
AAin
Output: X = 1 or 2; 1 = RS232, 2 = RS485
Laser targeting light:
AAlaX
X = 0 switch off laser targeting light
X = 1 switch on laser targeting light
Lock keyboard:
AAlkX
X = 0 ... 3
1 = lock lk1, removal with command lk0 or power
off-on
0 = removal of lock lk1
3 = continuous lock lk3, removal only with command
lk2
2 = removal of lock lk3