LSI LASTEM E-Log – User’s manual
52
Switch the display off manually or program it to automatically shut off by itself (see §4.3.4);
Disable the fast acquisition mode of the measures which was possibly enabled while the
instrument was switched on (see §5.2.5).
Cancel not-used actuation logics (see §5.5);
Manage modem switch on through actuator 7 (programmed with timed logic), or
programming GPRS communication through modem activation (see §3.1.1);
In case of sensors or telephone modems fed continuously, don’t use instrument’s batteries,
but use other ones (if possible). In this way the instrument can operate even if modem and/or
sensors aren’t fed, and so they aren’t operating;
Disable communication protocol on serial port 2, in case it isn’t used or in case it
corresponds to following types: TTY, Modbus, Gill anemometer.
ATTENTION
: Default configuration programmed on instrument is not the one of lowest energetic
consumption.
In case of Slave devices:
The Slave devices must have a channel to measure battery (level battery or voltage) that
bypasses the internal acquisition mode, normally set at one minute. The low energetic
consumption is achieved by programming the measure with a higher acquisition rate (5
minutes or more recommended).
It’s important don’t turn off (accidentally or otherwise programmed) the device “parent” of
the Slave because before the next attempt to send data, the Slave should enter into a search
mode (which still has a limitated duration) that consumes a lot of energy.
Содержание E-Log
Страница 1: ...Cod INSTUM_00351_en E Log User s manual Updated on 09 18 2017...
Страница 15: ...LSI LASTEM E Log User s manual 15...
Страница 60: ...LSI LASTEM E Log User s manual 60 6 6 Mask of terminal board...
Страница 61: ...LSI LASTEM E Log User s manual 61 6 7 Connection cables...
Страница 62: ...LSI LASTEM E Log User s manual 62 6 8 Declaration of Conformity...