LSI LASTEM E-Log – User’s manual
25
To shut off the display you can use the keyboard (see §4.2) or use program
3DOM
to have it shut
off; to do so, modify parameter "Characteristics – auto display shut off". By setting this parameter
to
Yes
, E-Log will shut off the display after the keyboard remains inactive for three minutes.
4.4.
LED's
The keyboard is fitted with two bicoloured LED's which show the operation of the instrument
(OK/ERR) and its communication with external devices in both directions (Rx/Tx).
Picture 8
See below a detailed description of each state.
Status indicator
Communication
indicator
Description
Off
-
Instrument switched off, or low feeding voltage
(min. 7 V)
Single quick green
blinking
-
Instrument switched on. Regular operation
(acquisition mode)
Three
quick
green
blinkings
-
Instrument switched on. The instrument
operates in reconfiguration mode
Five
quick
green
blinkings
-
The instrument is on. The instrument operates
in quick acquisition mode
Red slow blinking
-
Instrument switched on. An error occurred. The
number of blinkings indicates the type of error
(see §6.3)
-
Off
No current communication
-
Green blinking
Instrument is receiving data from the serial
communication lines
-
Red blinking
Instrument is transmitting data with the serial
communication lines
Interval between blinking cycles: 5 s.
Quick blinking period: 125 ms
Slow blinking period: 375 ms.
Содержание E-Log
Страница 1: ...Cod INSTUM_00351_en E Log User s manual Updated on 09 18 2017...
Страница 15: ...LSI LASTEM E Log User s manual 15...
Страница 60: ...LSI LASTEM E Log User s manual 60 6 6 Mask of terminal board...
Страница 61: ...LSI LASTEM E Log User s manual 61 6 7 Connection cables...
Страница 62: ...LSI LASTEM E Log User s manual 62 6 8 Declaration of Conformity...