Terminal Definition of Standard External Control Module

Analog  input  in  4 

20  mA,  0  -  5  V,  0  -  10  V  external  control 

modules. Control the speed of the pump.

External  control  star t/stop  input.  When  connected  to  COM,  the 
pump runs; When connected to high level, the pump stops.

External control cw/ccw input. When connected to COM, the pump 
rotates clockwise; When connected to high level, the pump rotates 
counter clockwise.

Analog  ground  (AGND)  in  4  -  20  mA,  0  -  5  V,  0  -  10  V  external 
control modules;
Pulse  input  in  pulse  input  external  control  modules.  Control  the 
speed of the pump. 10 KHz is corresponding to the max.speed.

The COM of external control direction and star t/stop signal input; 
It's  also  the  COM  of  pulse  signal  input  in  pulse  input  external 
control modules.

Terminal 1    

Terminal 2   

Terminal 3     

Terminal 4

Terminal 5

Pump is equipped with output por t in order to monitor the status of pump. 
Please see below “External Control Output Por t Pins Diagram: DB9”. Output 
signal adopts optoelectronic isolation circuit. A pull-up resistor and power 
supply are needed when using. 

Pins Definition

External Control Output

External Control Output Por t Pin Diagram: DB9

Footswitch Function

Connect Footswitch to r

 control Input Por t to control Star t/Stop of pump. 

emote

It can be set according to actual application. 

Pin 1   

Pin 2  

Pin 8  

Pin 4, 6, 7

Star t/Stop output. Pump outputs low level when it runs; pump 
output high level when it stops.

Direction  output.  Pump  output  low  level  when  it  runs  clock  wise; 
pump output high level when it runs count clock wise.

Pulse output, 0 - 600 rpm corresponding to 0 - 10 kHz.

COM

Footswitch is optional accessories. It can only control the star t and stop of the 
pump. The speed and direction of the pump are controlled by membrane keypad 
and knob on the operation penal.

Connection Diagram for Footswitch

Shor t terminal 5 and 11; shor t 
terminal 12 and 13; terminal 8 
connects footswitch; 
terminal 11 connects footswitch.

 

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

4 - 20 mA, 0 - 5V, 0 - 10V analog 
input control the speed

0 - 5V, 0 - 10V, 4 - 20mA

External Control Input Module Connection Diagram

+

-

1

4

CC/CCW Signal 
(5V TTL)

Star t/Stop Signal
(5V TTL)

+

-

3

5

+

-

2

0 - 10 kHz Pulse Signal Input Connection Diagram

0 - 10 kHz pulse input control 
the speed

+

-

4

5

CC/CCW Signal 
(5V TTL)

Star t/Stop Signal
(5V TTL)

+

-

3

5

+

-

2

WT600-1F OPERATING MANUAL

WT600-1F OPERATING MANUAL

Содержание WT600-1F

Страница 1: ...WT600 1F Peristaltic Pump Baoding Longer Precision Pump Co Ltd WT600 1F PERISTALTIC PUMP OPERATING MANUAL...

Страница 2: ...ds to be maintained or repaired Please shut down the power supply before install the external control equipments WT600 1F OPERATING MANUAL Table of Contents Instruction Operation Panel Basic Operation...

Страница 3: ...Direction Key Prime Key Knob LCD Display Power Switch Basic Operation Start Stop Key Press the Start Stop Key to start or stop the pump Direction Key Press the Direction Key to change the rotation dir...

Страница 4: ...t Key to switch these two interfaces a Current flow rates Display current flow rates Turn the Knob to increase or decrease the flow rates b Current speed Display current speed The speed will change wh...

Страница 5: ...2 Control Mode 2 Ext Cont ON 3 Address 1 System Setting All the setting can not be done when pump is running Pump head and tubing setting Press the Knob to highlight Pump Head Tube line press Knob ag...

Страница 6: ...previous menu Flowrate Dispensing Disp Vol 2 02mL Copy No 10 Disp Vol 2 02 mL Copy No 10 1 System Setting 2 Control Mode 1 System Setting 2 Control Mode 3 Disp Setting 4 Calibration 1 System Setting 2...

Страница 7: ...ob to select test time its range is from 30 to 1800 seconds press Knob for confirmation Press Exit Key to return to previous menu Press Start Stop Key the LCD displays running time and actual volume R...

Страница 8: ...nd tubing refer to the table of page 4 Press Knob to enter System Setting interface to set pump head and tubing Set Back Suction rpm Enter control mode interface to select Dispensing Enter Dispense Se...

Страница 9: ...ns Definition External Control Output External Control Output Port Pin Diagram DB9 Footswitch Function Connect Footswitch to r control Input Port to control Start Stop of pump emote It can be set acco...

Страница 10: ...15X YZ2515X YZII15 YZII25 DG15 24 KZ25 BZ25 DMD25 Operation Mode Membrane keypad and Knob Direction Control CW and CCW reversible Prime Full speed for fast filling and emptying Back Suction Prevent th...

Отзывы: