6

5

Menu Diagram

Changing the Pump Head

The pump head is mounted on drive before the pump leaves the factory. Follow 
below procedure for changing another pump head.

Loose  the  two  M4  screws  which  connect  the  pump  head  and  the  drive. 
Dismantle the pump head slightly.
Inser t the flat end of the pump head's shaft to the slot of drive's coupling. 
Make the location hole of the pump head match the location pin of the drive.
Inser t the two mounting screws into the mounting holes of the pump head. 
Then  tighten  the  mounting  screws  to  connecting  hole  of  drive.  (For  more 
information, please refer to pump head's operation manual)

 

Turn off the power supply before changing the pump head.

1.

2.

3.

VOL

:  

0.25mL

Copy

: 

991

Running Inter face

1.System Setting

2.Control Mode

3.Disp Setting

4.Calibration

1. Pump head

&

Tube

YZ1515    ID 0.80

2.Ext  Cont. 

ON

3.Address 

:  

1

4.Footswitch 

G

5.Back  rpm 

1

.

2

"

Flowrate

Dispensing

Disp Vol

:

2.02mL

Copy No. 

10

Flow

:

2.043mL/min

Pause rpm 

1.0 

"

Flowrate
Dispensing

1.System Setting

2.Control Mode

3.Disp Setting

4.Calibration

 

YZ1515  ID 0.80

YZ1515

YZ2515

Select

 

Tubing  ID

0.80   1.60   2.40

Ext  Control 

 

ON

    

OFF

Pump Address

    

1

1.System Setting

2.Control Mode

1. Pump head

&

Tube

1.System Setting

2.Control Mode

2.Ext  Cont. 

ON

3.Address 

:  

1

1.System Setting

2.Control Mode

2.Ext  Cont. 

ON

3.Address 

:  

1

System Setting:

All the setting can not be done when pump is running.

Pump head and tubing setting

Press the 

Knob

 to highlight Pump Head & Tube line, press 

Knob

 again to enter 

pump  head  selection  inter face,  turn 

Knob

  to  select  the  pump  head,  press  the 

Knob

 for confirmation and enter tubing selection inter face, turn 

Knob

 to select 

the  tubing,  press 

Knob

  for  confirmation  and  return  to  pump  head  selection 

inter face. Press 

Exit Key

 to return to previous menu.  

External Control setting

Set the pump to enable the external control or not.

On: Enable external control (0 - 5 V, 0 - 10 V, 4 - 20 mA, 0 - 10 KHz)

Off: Disable

 

external control

Set the external control

Press  and  turn 

Knob

  to  highlight  the  external  control  line.  Press 

Knob

  to  enter 

next inter face, turn 

Knob

 to select On or Off, press 

Knob

 for confirmation. Press 

Exit Key

 to return to previous menu.

Press and turn 

Knob

 

Press and turn 

Knob

 

Address setting:

When control computer controls many pumps through RS485 inter face, it must 
identify each pump's I.D. This pump I.D. should be unique. It's the identification 
of this pump. Maximum 30 pieces WT600-1F can be controlled through RS485 
at the same time.

Setting the address

Press and turn 

Knob

 to

 

highlight the address line. Press 

Knob 

again to enter next 

inter face, turn 

Knob

 to select Pump I.D. number, press 

Knob

 for confirmation. 

Press 

Exit Key 

to return to previous menu.

WT600-1F OPERATING MANUAL

WT600-1F OPERATING MANUAL

Содержание WT600-1F

Страница 1: ...WT600 1F Peristaltic Pump Baoding Longer Precision Pump Co Ltd WT600 1F PERISTALTIC PUMP OPERATING MANUAL...

Страница 2: ...ds to be maintained or repaired Please shut down the power supply before install the external control equipments WT600 1F OPERATING MANUAL Table of Contents Instruction Operation Panel Basic Operation...

Страница 3: ...Direction Key Prime Key Knob LCD Display Power Switch Basic Operation Start Stop Key Press the Start Stop Key to start or stop the pump Direction Key Press the Direction Key to change the rotation dir...

Страница 4: ...t Key to switch these two interfaces a Current flow rates Display current flow rates Turn the Knob to increase or decrease the flow rates b Current speed Display current speed The speed will change wh...

Страница 5: ...2 Control Mode 2 Ext Cont ON 3 Address 1 System Setting All the setting can not be done when pump is running Pump head and tubing setting Press the Knob to highlight Pump Head Tube line press Knob ag...

Страница 6: ...previous menu Flowrate Dispensing Disp Vol 2 02mL Copy No 10 Disp Vol 2 02 mL Copy No 10 1 System Setting 2 Control Mode 1 System Setting 2 Control Mode 3 Disp Setting 4 Calibration 1 System Setting 2...

Страница 7: ...ob to select test time its range is from 30 to 1800 seconds press Knob for confirmation Press Exit Key to return to previous menu Press Start Stop Key the LCD displays running time and actual volume R...

Страница 8: ...nd tubing refer to the table of page 4 Press Knob to enter System Setting interface to set pump head and tubing Set Back Suction rpm Enter control mode interface to select Dispensing Enter Dispense Se...

Страница 9: ...ns Definition External Control Output External Control Output Port Pin Diagram DB9 Footswitch Function Connect Footswitch to r control Input Port to control Start Stop of pump emote It can be set acco...

Страница 10: ...15X YZ2515X YZII15 YZII25 DG15 24 KZ25 BZ25 DMD25 Operation Mode Membrane keypad and Knob Direction Control CW and CCW reversible Prime Full speed for fast filling and emptying Back Suction Prevent th...

Отзывы: