11
12
Select pump head
and tubing refer to
the table of page 4
Dispensing mode operation procedure
Confirm the
volume to be
dispensed
Press
Exit Key
to
return to running
inter face
Operation Procedure
Before running the pump, select pump head and tubing refer to “Acceptable
Pump Head and Tubing (Table 1)”. Thick wall tubing is preferred for longer
service life.
Flow rates mode operation procedure
Confirm the flow
rates of the fluid
which need to be
pumped
Press
Knob
to
enter
System
Setting
inter face
to set pump head
and tubing
Enter
Control
Mode
inter face to
select
Flow Mode
Press
Exit Key
to
return to running
inter face, turn
Knob
to select
needed flow rates
Press
Knob
to
enter calibration
selection inter face
and select
Flowrate
when
calibration is
needed
Set
Test Time
,
press
Start/Stop
to star t the pump.
Input the actual
volume
Press
Exit Key
to
return to running
inter face
Select pump
head and tubing
refer to the table
of page 4
Press
Knob
to
enter
System
Setting
inter face
to set pump head
and tubing
Set
Back
Suction rpm
Enter control
mode inter face
to select
Dispensing
Enter
Dispense
Setting
inter face
to set all the
dispensing
parameters.
Press
Knob
to
enter calibration
selection
inter face and
select
Dispensing
when calibration
is needed
Press
Start/Stop
to star t the pump.
Input the actual
volume
External Control Input
Analog Signal Input function
Communication Control
The pump can connect to control computer (computer, PLC, SCM) through
RS485 serial communication module (shown as below). Please contact Longer
Company for communication protocol.
COM
B
A
Set the “External Control” in “on” state. External control module is shown as below.
Five kinds of standard external control module need to be ordered separately
according to special requirements.
5
4
3
2
1
External Control Module
Analog Control
4 - 20 mA Speed Control
0 - 5 V Speed Control
0 - 10 V Speed Control
0 - 10 KHz Speed Control
External Control
DB15
Communication Control RS485
WT600-1F OPERATING MANUAL
WT600-1F OPERATING MANUAL