ACS-CB-SENS - MANUAL // PAGE 15
6.6 CONNECTING THE LOG-PRT-DMS TO THE ACS-CB-
SENS
NOTICE
To avoid strain on Cables, ensure the LOG-PRT-DMS is mounted a suitable distance from
the ACS-CB-SENS.
DANGER
The RJ sockets on the ACS-CB-SENS may only be used to accommodate LOGICDATA-
approved sensor units. Inserting other telecommunication-type sensors into these
terminals may cause damage to the Control Box or other products in the system.
For telecommunication type connectors not used for connection to the telecommuni-
cation network should be provided with a marking identifying the specific function or
circuit characteristic the connector or terminal is used for.
INFO
If your system is not parametrized to support Intelligent System Protection, you will
need to alter the parameters so that the LOG-PRT-DMS Collision Sensor functions cor-
rectly. You may also need to adjust parameters if extra sensor units are added to the
system. Contact LOGICDATA for further advice on system parameters.
1. Ensure the ACS-CB-SENS is not connected to the Control Box.
2. Insert the RJ10 connector(s) of the LOG-PRT-DMS Sensor(s) into the Sensor Port(s).
Fig. 5:
Connecting the LOG-PRT-DMS to the ACS-CB-SENS
NOTICE
In Cascading Systems, only one LOG-PRT-DMS sensor may be connected
NOTICE
If only one LOG-PRT-DMS is used, connect it to the plug port marked "1".
RJ10 Connector