18
5.3
Changing the Syringe
The unit need not be switched off to change the syringe.
●
Press the hose clamp
11
closed so that pressure cannot be applied to the syringe adapter
12
as a result of an unintentional start.
●
Replace the Luer-Lock tip cap
17
on the filled syringe
15
with a dispensing needle
16
.
●
Replace the syringe and release the hose clamp
11
.
☞
To avoid air bubbles during dispensing, the tip of the syringe must be purged of air.
●
Hold the syringe
15
over a container
since product will flow out!
●
Press button
to switch to
cont
inuous mode
(see Section 5.2.2.2).
●
Press and hold button
(or the
Foot Switch
20
) until the product flows free
of bubbles from the dispensing needle
16
.
●
Press button
to switch off
cont
inuous mode.
●
Continue dispensing with the stored values or
readjust the dispense quantity according to Section 5.2.2.
For a change in the type of product,
readjust the dispense quantity according to Section 5.2.2.