8-3 Short-Spin 

1) Please set speed using the button 

  (RPM/RCF) and 

  ( , ). 

2) The unit stops after reaching at preset RPM. 
 
 

8-4 Opening the centrifuge in the event of power failure 
 

  In case of a power failure during operation, the lid cannot be opened. 

If power is supplied again the alarm will beep and the rotor will stop. 
The lid cannot be opened until the rotor comes to a complete stop. 

 

  Please make sure that the rotor has come to a complete stop. 

Lift up Mini Centrifuge, insert a pen into the opening in the base and move the 
lever in the direction of the arrow. 

 

  When lifting up the centrifuge, particularly after a device failure, please 

note that the motor on the underside of the centrifuge can become hot. Please 
hold the device at the edges only. 

 
 

9. Maintenance 

Cleaning 

 

The  rotor  and  outside  of  the  centrifuge  should  be  cleaned  regularly  with  a  damp 
cloth. 

 

Disconnect the centrifuge from the main supply before the cleaning. 

 

Only neutral agents or a mild liquid soap may be used. 

 

Do not use chemical cleaning agents or strong solvents. 

 

 

Mechanical or chemical damage 

 

Do not operate the centrifuge when there is mechanical or chemical damage. 

 

Change faulty parts immediately. 

 

 

15

DOMINIQUE DUTSCHER SAS

Содержание MCF-1350

Страница 1: ...OPERATIONS MANUAL Mini Centrifuge Model MCF 1350 Certified D O M I N I Q U E D U T S C H E R S A S...

Страница 2: ...Specifications 8 5 Features 9 6 Parts of the Mini Centrifuge 10 7 Control elements and Function 11 8 Installation and Operation 13 9 Maintenance 15 10 Trouble shooting 16 11 Technical data 17 12 Warr...

Страница 3: ...an inappropriate power supply may result in damage to the equipment Adhere to all electrical codes This device must be connected to an earthed power supply Insert the plug securely into a wall socket...

Страница 4: ...evice near a heat source or where temperatures are elevated Do not operate the device when temperatures fall below 4o C and rise above 45o C Do not use this device where there are flammable or corrosi...

Страница 5: ...warm location operate the device for 30 minutes prior to the move to avoid excessive condensation Alternatively allow it to warm to room temperature for at least three hours before plugging the socket...

Страница 6: ...operation Do not use the device to process samples with flammable contents This unit is not designed to be explosion proof Do not expose the device to Ultra Violet UV light The lid may craze and crac...

Страница 7: ...please pay attention to the alert symbols throughout the manual 3 Package Contents Mini Centrifuge 1 unit Angle rotor for 1 5 2 0ml micro tubes 1 ea Lid for Angle rotor 1 ea Nut for Rotor Fixing 1 ea...

Страница 8: ...or rotor Stainless Steel Autoclavable Material Nut for Rotor Stainless Steel Autoclavable Timer 1 30 min Noise Level Less than 60dB Acceleration Time 12 sec Braking Time 16sec Dimension mm 208 W x 245...

Страница 9: ...rifugal force is 12 300 x g Easy to follow digital display Separate short spin button Max temp in chamber at Max rpm 35 at room temp 24 Automatic lid release Maintenance free drive Autoclavable rotor...

Страница 10: ...6 Parts of the Mini Centrifuge Front View Rear View 10 D O M I N I Q U E D U T S C H E R S A S...

Страница 11: ...yed figure x1000 Actual rpm or rcf RPM Display It is displayed when you select RPM operation mode RCF Display It is displayed when you select RCF operation mode Operation Display It is displayed when...

Страница 12: ...to show RPM or RCF Door To open the lid Opening the centrifuge in the event of power failure Please make sure whether the rotor has come to standstill Then lift up Mini Centrifuge insert a pen into th...

Страница 13: ...de of device 3 Switch ON the power switch on the rear The door is opened automatically and the centrifuge is ready to operate when the display becomes visible 4 Place the rotor onto the rotor axle and...

Страница 14: ...re x1000 Actual rpm or rcf 2 Setting Timer 1 Please push the Time Button 2 Please set the time using the button and The timer can set between 1 30 min 3 Please push the Start Stop button In Operation...

Страница 15: ...and move the lever in the direction of the arrow When lifting up the centrifuge particularly after a device failure please note that the motor on the underside of the centrifuge can become hot Please...

Страница 16: ...load the tubes symmetrically Ensure that the tubes on the opposite sides are equally filled Excessive Noise or Vibration Rotor is slanted Remove the rotor and reinsert straight and firmly downwards E1...

Страница 17: ...es Max Centrifugal force 12 300 rcf Rotor for 1 5 2 0 ml micro tubes Ambient temperature 2 35 o C Max relative humidity 75 non condensing moisture Degree of pollution N A Overvoltage category N A Max...

Страница 18: ...nged without charge Warranty is void on devices with damaged manufacturer s number damaged seal or damage due to improper use storage or transport of device or service by unauthorized persons CERTIFIC...

Отзывы: