8. Installation and Operation
8-1 Installation
1) Install your Mini Centrifuge on a clear, flat and level surface.
CAUTION
Place the centrifuge on a level, horizontal surface. Please make sure that
the ventilation slits are not blocked. IEC 1010-2-020 stipulates that 30 cm
around the centrifuge should clear of all objects and obstructions.
2) Connect the power cord to the main power socket on the rear side of device.
3) Switch ON the power switch on the rear. The door is opened automatically and
the centrifuge is ready to operate when the display becomes visible.
4) Place the rotor onto the rotor axle and tighten using the rotor nut.
5) Place tubes symmetrically into the rotor
CAUTION
- Unbalanced loading will significantly reduce the life of your centrifuge.
- Load your tubes symmetrically to ensure safe and quiet operation.
- To spin1 or 5 tubes, load an additional water filled tube in the opposite slot.
6) Close the lid for Angle rotor
7) Close the centrifuge lid.
13
DOMINIQUE DUTSCHER SAS