8
ENG
USE OF THE DEVICE
Important!
Before using the device for the
fi
rst time it is necessary
to make a full cleaning, as described in item 1 of section «Care,
storage, repair and recycling»
1. Insert the inhalation chamber into the slots at the top of the main
unit until it clicks (Fig. 4).
Attention!
Keep the electrodes of the device and inhalation
chamber clean otherwise nebulizer may not work.
Important!
The service life of a mesh membrane of inhalation
chamber is about six months when using three times a day.
2. Open the cover of of inhalation chamber, lifting the latch (Figure 5). Fill the solu-
tion into the medicinal compartment (Figure 6). The maximum capacity of the
compartment for drugs – 8 ml (up to the mark MAXLEVEL), minimum – 0.5 ml.
In order to maintain the temperature of the warm solution, complete compart-
ment for water with hot water.(Fig. 6) It is recommended to
fi
ll the water heated
to 80°C. Close the cover of inhalation chamber, lowering the latch.
Do not use excessive force when opening and closing the latch (This will
increase the life of an inhalation chamber).
Make sure the lid is closed tightly, and inhalation solution will not pour out
from the inhalation chamber.
3. Place the adapter on the mouthpiece inhalation chamber, while tabs on the
inhalation chamber should fall into the holes on the adapter.
4. Place the mouthpiece (Figure 7) or mask (via an adapter for masks) (Figure 8).
5. To start the procedure of inhalation, touch the key
POWER indicator will turn green.
Attention!
If inhalation chamber is empty when power is on, the POWER indicator
will illuminate for one second, then the device automatically turn off.
It is possible that after the power is on, the nebulizer will spray the
inhalation solution for one second, then pauses for 0,5 seconds. It is
the normal operation of the device, however, after this pause, the device
must spray inhalation solution continuously.
INHALATION
For effective inhalation need to calm down, relax and sit up straight If you make
inhalation in bed, then put a pillow under your back so that the back was as straight
as possible Inconvenient posture and irregular breathing can lead to blockage the part of the lung and
certain areas of the respiratory tract.
Important!
Each patient is recommended to use an individual mouthpiece and / or mask.
Take a comfortable position and start inhalation. Breathing should be smooth, breathe slowly and
deeply to spray deep into the respiratory tract.
Fig.6
Medicine
compartment
Water
compartment
8ml
Fig.4
8ml
Fig.5