• Never plug in or unplug from the power socket with wet hands.
• Do not immerse the main unit into water or other liquids.
• Protect it against water or other liquid splashes. In case it contacts with any splash, disconnect the
power supply and mop up with gauze or other soft adsorbing materials.
• Never use the device in humid environment such as bathroom.
• When connected to power supply, check the supply voltage at power socket is correct and make
sure your plugging in will not cause overload.
• Disconnect it from power supply after the use of this device and never leave an energized device aside.
• Clean the device only after disconnecting it from power supply.
INSTALLATION OF BATTERIES
The set of nebulizer includes two AA batteries (LR6). Included batteries are
designed to verify the operability of the device, and their service life may be
shorter than service life of new batteries.
To install batteries, open the battery compartment cover by pressing a
fi
nger on the
latch cover (Figure 1). Insert the batteries observing the polarity (Figure 2) Close
the battery compartment cover by pushing your
fi
nger until click. New batteries,
depending on the type and capacity, can provide work an average of 4 days (when
using for 20 minutes per day).
When the WATT indicator is
fl
ashing, the nebulizer may work for another 10
minutes. If the WATT lights continuously, it means that the level of charging is too
low. Nebulizer stops working. Replace all batteries with new ones.
• Do not leave used batteries inside the device.
• To power the device recommend purchasing alkaline batteries type AA (LR6)
or rechargeable Ni-MH type.
• Do not use saline batteries such as R6.
USE THE DEVICE WITH POWER SUPPLY
Use a power supply (sold separately) with the following specifications:
Output voltage
: 3V ± 5%
Current consumption : at least 500 m
А
Plug:
Polarity
:
“minus” – the internal contact
Outside diameter:
: 3,5 ± 0,1 mm
Inner diameter
: 2,1 ± 0,1 mm
Length
: 9 ± 0,3 mm
Slot for the source: is located on the back side of the device (Fig. 3)
To use the device with power supply, connect the plug to the power supply unit and
plug the power supply plug into the power outlet and touch the key . Finished the
measurement turn off the device by touching the key , unplug the power supply
from the wall outlet and disconnect the power supply connector from the device.
Fig.2
Fig.3