D9S001GE4-0101
43/
57
Scale Calibration
–iBoard Basic version
5.3.1
Introduction
5.3.1.1
EU/US version
The display shows the load value. This can be adjusted to the required value by using the
ZERO/T
or
HOLD
functions. If the value is negative, the first segment shows a minus sign. If the value is lower than -5.0kg (11.0 lb),
the display shows “
Lo
”. If the value is higher than 254,5 kg (561,1 lb), the display shows “
Hi
”.
If the value of absolute weight is lower than 0, the display shows “
– – Lo
”.
5.3.2
Diagnostic mode
In this mode it is possible to browse the scale system values.
Scale service mode activation:
Press the
ZERO
,
CLEAR
and
HOLD
buttons simultaneously for 5 seconds.
To exit diagnostic mode press CLEAR button
5.3.2.1
General instructions
Display:
The identification number and respective value is displayed
The first value displayed is always position
TENS 1:
Identification
number
Value meaning
Value
TENS 1
Absolute un-recalculated value
– tensometer 1 (rounded to ten)
TENS 2
Absolute un-recalculated value
– tensometer 2 (rounded to ten)
TENS 3
Absolute un-recalculated value
– tensometer 3 (rounded to ten)
TENS 4
Absolute un-recalculated value
– tensometer 4 (rounded to ten)
ABS WGH
Absolute weight
ABS 0
Absolute 0
CON 1
Constant
– Tensometer 1
1,700 - 2,200
CON 2
Constant
– Tensometer 2
1,700 - 2,200
CON 3
Constant
– Tensometer 3
1,700 - 2,200
CON 4
Constant
– Tensometer 4
1,700 - 2,200
CON GR
Gravitation constant
1 - 8
CON CAL
Calibration constant
1,000
NR CAL
Number of calibrations
Содержание Wissner-Bosserhoff ELEGANZA 4
Страница 5: ...D9S001GE4 0101 5 57 Identification of bed sides...
Страница 18: ...D9S001GE4 0101 18 57 Weight module and tensometers S6015189ND...
Страница 31: ...D9S001GE4 0101 31 57 i Drive Power connection scheme...
Страница 34: ...D9S001GE4 0101 34 57 Castors Castors Antistatic castor Castor with lock Central Castor Control...
Страница 35: ...D9S001GE4 0101 35 57 Bed Wiring Diagram...
Страница 45: ...D9S001GE4 0101 45 57 The CAL position displays the calibration constant...
Страница 49: ...D9S001GE4 0101 49 57 5 3 6 4 Positioning the weights for sensors 2 head left sensor...
Страница 51: ...D9S001GE4 0101 51 57 5 3 6 6 Positioning the weights for sensor 3 left foot sensor...
Страница 53: ...D9S001GE4 0101 53 57 5 3 6 8 Positioning the weights for sensor 4 right foot sensor...