D9S001GE4-0101
42/
57
5 Bed Setup
Column and accelerometer calibration
After replacing always calibrate:
a) Any column.
b) Control box.
c) Accelerometer (iBoard Basic)
d) Any iBoard Basic
Calibration of columns means:
Calibrating the speed of the columns during up or down movement.
Calibration of accelerometers means:
Calibrating the mattress platform, backrest and calfrest accelerometers in the horizontal position.
5.1.1
Column and accelerometer calibration using the ACP
Ensure the bed is on a level floor!
Enter the calibration mode by pressing a combination of four buttons on the nurse control panel.
Simultaneously press backrest up, backrest down, CPR and anti-Trendelenburg buttons, and hold for 3 seconds.
Check whether you are in calibration mode. The lock functions and battery LEDs should be flashing on the ACP.
Control panel as well as the battery LED on the iBoard panel and the lock function LEDs on the hand control.
This mode is only active for 1 minute.
Hold CPR button. The bed will go UP, DOWN and UP. After each end position one LED lock will turn off.
In case of unpress of the CPR button during the calibration, the calibration stop and must be repeat.
Transport lock
Simultaneously press the
calfrest down
,
calfrest up, and Trendelenburg buttons
, and hold them for 3
seconds to activate the transport lock. The bed must be disconnected from the mains.
Содержание Wissner-Bosserhoff ELEGANZA 4
Страница 5: ...D9S001GE4 0101 5 57 Identification of bed sides...
Страница 18: ...D9S001GE4 0101 18 57 Weight module and tensometers S6015189ND...
Страница 31: ...D9S001GE4 0101 31 57 i Drive Power connection scheme...
Страница 34: ...D9S001GE4 0101 34 57 Castors Castors Antistatic castor Castor with lock Central Castor Control...
Страница 35: ...D9S001GE4 0101 35 57 Bed Wiring Diagram...
Страница 45: ...D9S001GE4 0101 45 57 The CAL position displays the calibration constant...
Страница 49: ...D9S001GE4 0101 49 57 5 3 6 4 Positioning the weights for sensors 2 head left sensor...
Страница 51: ...D9S001GE4 0101 51 57 5 3 6 6 Positioning the weights for sensor 3 left foot sensor...
Страница 53: ...D9S001GE4 0101 53 57 5 3 6 8 Positioning the weights for sensor 4 right foot sensor...