LINET Wissner-Bosserhoff ELEGANZA 4 Скачать руководство пользователя страница 12

D9S001GE4-0101 

 
 

 

 

12/

57

 

 

 

 

Head side rails 

The head side rails are at the sides of the head section of the bed's mattress platform rear section. 

3.3.1 

iBoard Basic 

The iBoard Basic control panels are a basic control elements. They provide for configuration of bed functions, and 

control, display of the weighing system and bed exit alarm. They are built into the outer side of both head side rails, 

to which they are connected using six screws covered by six adhesive blanking plugs. They are connected using 6-

core cables with a mini Din connector to the cable in the side rail mechanism (It leads from the control unit). 

The iBoard Basic is equipped with 

2,4”

 display and keyboard with buttons. There are no positioning buttons on the 

panel. On the rear side of the panel is a production and technical label, and the backrest accelerometer. 

Due to absence of STOP button the bed must be equipped with ACP with STOP button. 

 

3.3.2 

Backrest accelerometer in iBoard Basic control panels 

The backrest accelerometer is a part of the both iBoards Basic and is used to determine the 

tilt

 of the bed's 

backrest. When the tilt of the backrest changes gradually due to gravity, the voltage in the accelerometer changes 

and the angle of tilt is calculated. This information is sent to the control unit, which then displays it on the panel. The 

angle of tilt is relative to the 

ground

, which means that if the bed is not on a horizontal floor the backrest of the bed 

is tilted, the angle of tilt of the floor (ground) is added to the angle of tilt of the frame. The control panels cannot be 

interchange 

– Left must be on the left side and right on the right side. In case of incorrect connection it displays 

wrong tilt angle of the backrest. 
 

 

Содержание Wissner-Bosserhoff ELEGANZA 4

Страница 1: ...D9S001GE4 0101 Version 03 Publication Date 2019 09 Service Manual ELEGANZA 4 Positionable Bed for Intensive Care version with scales and without scales...

Страница 2: ...ers in foot side rails 13 Supervisor control panel ACP 15 Handset 15 Mobi Lift 15 3 7 1 Left and right side rail limit switches EMR ready Only 16 Weighing module 16 Weighing sensor tensometer 17 Integ...

Страница 3: ...play 39 4 3 2 Stop codes 40 4 3 3 Operation codes 41 5 Bed Setup 42 Column and accelerometer calibration 42 5 1 1 Column and accelerometer calibration using the ACP 42 Transport lock 42 Scale Calibrat...

Страница 4: ...echnical label description 1 Address of Distributor 2 Manufacturing Date Year Month Day 3 DI Device Identifier GTIN Global Trade Item Number 4 1D Bar code GS1 128 Serial Number 5 Symbols 6 Configurati...

Страница 5: ...D9S001GE4 0101 5 57 Identification of bed sides...

Страница 6: ...D9S001GE4 0101 6 57 2 Technical data and Symbols Follow User Manual for Eleganza 4 D9U001GE4 0101 to find Technical Data and used Symbols...

Страница 7: ...and sends it to the control unit Mains power is supplied to the power box via a three core mains cable in the relevant national standard The power unit is standard at voltage 230V 50 60 Hz and altern...

Страница 8: ...he leak is recorded simultaneously in the permanent memory of the control unit Positioning function locks and safety circuit to detect overload See Detecting overload 3 1 4 Technical parameters Parame...

Страница 9: ...ccidental movement of the bed during a fault or from unintentional pressing of controllers 3 1 5 4 Simultaneous pressing of any two buttons The control unit is issued with the function for recognizing...

Страница 10: ...ndition After an additional increase of approximately 15 C the switching power elements are disconnected this makes positioning impossible also signalled by a constant tone Information about the therm...

Страница 11: ...e mode Information about charging and modes of the rechargeable battery is signalled by the rechargeable battery s LED as follows Off The rechargeable battery is charged one exception may be a certain...

Страница 12: ...e panel is a production and technical label and the backrest accelerometer Due to absence of STOP button the bed must be equipped with ACP with STOP button 3 3 2 Backrest accelerometer in iBoard Basic...

Страница 13: ...the control unit indicator lights for the mode and condition of the rechargeable battery a selection of special functions and settings for standard positions of the bed Foot siderails 3 4 1 Controller...

Страница 14: ...D9S001GE4 0101 14 57 Number Description 1 iBoard Basic Cable Left Right 2 Foot Siderail Control Cable Siderail Cables 2 1 2 2 1 1...

Страница 15: ...functions of both handsets are identical A cable with an 8 pin DIN connector is connected to a Plug Play adaptor which is linked to the control unit The controller has an RC unit to detect short circu...

Страница 16: ...e side rail down Weighing module The weighing module allows the bed s mattress platform load to be calculated Four weighing sensors tensometers are connected to the weighing module The tensometer cons...

Страница 17: ...lded cable The functions of the individual cores are described in the table Model number of tensometer 8300 0674A Model number of tensometer 8300 0674A Sensors T1 T2 PIN Meaning Colour Sensors T3 T4 P...

Страница 18: ...D9S001GE4 0101 18 57 Weight module and tensometers S6015189ND...

Страница 19: ...ls to the server Each device is marked with a label containing important information about the integration module The label contains the product number of the product power supply level type of hardwa...

Страница 20: ...D9S001GE4 0101 20 57 EMR ready solution Mattress platform actuators Position Description 1 Backrest linear actuator 2 Thighrest linear actuator Mattress platform actuators 1 2...

Страница 21: ...ble 1435 mm Compression strength 3 00 kN Tensile strength 0 kN Mechanical release Yes Input voltage 24 V Overvoltage protection No Cover IPX6 Max current load 3 A 3 12 2 Thighrest actuator The thigh m...

Страница 22: ...olex connector data undercarriage module The head column is located near back rest part and foot column is located under the leg rest area The columns are fixed using two clamps to the frame undercarr...

Страница 23: ...castor I Brake Voltage is measurable at the connector on pin number 1 and 3 see diagram connector No 8 The analogue signal of the UM allows the presence of mains power to the bed to be identified Thi...

Страница 24: ...rriage frame with two screws 11 1 Nm There are three buttons on the controller lift upwards lift downwards and lift to examination position A cable runs from the foot controller to the undercarriage m...

Страница 25: ...trol CCC replacing the activity of the operator The position of the CCC braked unbraked is detected by a single magnetic limit switch Its signal is transmitted via the undercarriage module to the cont...

Страница 26: ...the bed from injuries due to crushing by the lowered Mattress Platform When obstacle occurs on the undercarriage and Mattress Platform is going down the motion is automatically stopped Safe Stop consi...

Страница 27: ...d If the i Drive has been fully discharged the functions of the bed can still be used The bed has one i Drive Power controller 3 20 1 Technical specifications Specifications Value Diameter of i Drive...

Страница 28: ...controller or the touch sensor is switched on The i Drive castor is controlled using the i Drive controller The controller contains four pressable buttons and one touch sensor If Forward and Reverse a...

Страница 29: ...t is equipped with an input for a 36V rechargeable battery an AC charging source an input for the Braked Bed magnetic switch and an analogue motor temperature sensor input When the bed is plugged into...

Страница 30: ...o that its lower voltage during gradual discharge is compensated for 3 20 8 i Drive rechargeable battery The driven castor rechargeable battery comprises three 12V rechargeable batteries connected in...

Страница 31: ...D9S001GE4 0101 31 57 i Drive Power connection scheme...

Страница 32: ...on in Control Unit Thermal Sensor Cable 13 Position in Control Unit Castor Control Position Cable with Sensor A i Brake Cable B i Drive Power Wheel Emergency Retraction Cable with switch C i Drive Pow...

Страница 33: ...on 2 5th castor foot controled 3 Distance washer Fe Zn Cr d6 4 D18 T1 6 DIN 9021 4 Screw Fe Zn Cr M6 L20 DIN 7500C TORX 5 screw Fe Zn M6 L20 DIN 912 VAL Hex 6 Nut M 6 DIN 985 7 spring D9 T0 8 L80 DIN...

Страница 34: ...D9S001GE4 0101 34 57 Castors Castors Antistatic castor Castor with lock Central Castor Control...

Страница 35: ...D9S001GE4 0101 35 57 Bed Wiring Diagram...

Страница 36: ...of electrical parts of bed Turns off after removed of the issue CU or press of the STOP button SAFE STOP Movement of the Mattress support platform stopped by function Safestop Remove an object from u...

Страница 37: ...tery discharged completely Only CPR Blinking 100ms ON 100ms OFF Battery charging Blinking 200ms ON 1600ms OFF NOTE First value always means blinking of the LED and the second value means the LED is tu...

Страница 38: ...faulty indicated on accumulator status indicator Flashes i Drive control box heat protection is activated 4 2 3 Fault signalisation i Drive Power The system is protected against failure states by stop...

Страница 39: ...celerometer 0 80 Chair 8 AD STOP AD stop 100 120 1x STOP 65 85 2x STOP 9 OPER C Operation codes see 4 3 3 XX YY XX function code YY code of periphery 10 STOP C Bed stop code see 4 3 2 11 SYS ERR Syste...

Страница 40: ...rature exceeded 7 column current overload 0 8 column current overload 1 10 actuator current overload backrest 11 actuator current overload thighrest 12 stopped due to resistance on the same function 1...

Страница 41: ...eight adjustment 9 Lock pedals 10 Thighrest up 12 Thighrest down 13 Trendelenburg 15 Backrest up 16 CPR 17 Backrest down 18 Anti Trendelenburg 19 Mobilisation position 20 Height adjustment up 21 Cardi...

Страница 42: ...nurse control panel Simultaneously press backrest up backrest down CPR and anti Trendelenburg buttons and hold for 3 seconds Check whether you are in calibration mode The lock functions and battery L...

Страница 43: ...simultaneously for 5 seconds To exit diagnostic mode press CLEAR button 5 3 2 1 General instructions Display The identification number and respective value is displayed The first value displayed is a...

Страница 44: ...esignated measuring instrument must be verified by an authorised person after any change of constants according to the metrological act 5 3 4 Calibration mode Press and hold the HOLD CLEAR ZERO and SP...

Страница 45: ...D9S001GE4 0101 45 57 The CAL position displays the calibration constant...

Страница 46: ...symbol L 0 flashes the on the display The bed must be balanced throughout the calibration In this phase is set absolute zero value 5 3 5 3 Phase Three L160 load 160 is displayed on the display It is p...

Страница 47: ...5 it needs to calculate with actual displayed value 5 3 6 1 Location of the tensometers on the bed No 1 Head left No 2 Head right No 3 Foot left No 4 Foot right Zeroing the scale Zero display while t...

Страница 48: ...D9S001GE4 0101 48 57 5 3 6 3 Calculating and saving the new sensor constant sensor 1 sensor 1 EU version Press the button 0 5 0 1 di Write the new constant value to the position of sensor 1 87 5...

Страница 49: ...D9S001GE4 0101 49 57 5 3 6 4 Positioning the weights for sensors 2 head left sensor...

Страница 50: ...D9S001GE4 0101 50 57 5 3 6 5 Calculating and saving the new sensor constant sensor 2 sensor 2 EU version Press the button 0 5 0 1 div Write the new constant value to the position of sensor 2 87 5...

Страница 51: ...D9S001GE4 0101 51 57 5 3 6 6 Positioning the weights for sensor 3 left foot sensor...

Страница 52: ...D9S001GE4 0101 52 57 5 3 6 7 Calculating and saving the new sensor constant sensor 3 sensor 3 EU version Press the button 0 5 0 1 Write the new constant value to the position of sensor 3 87 5...

Страница 53: ...D9S001GE4 0101 53 57 5 3 6 8 Positioning the weights for sensor 4 right foot sensor...

Страница 54: ...D9S001GE4 0101 54 57 5 3 6 9 Calculating and saving the new sensor constant sensor 4 sensor 4 EU version Press the button 0 5 0 1 Write the new constant value to the position of sensor 4 87 5...

Страница 55: ...bration mode The calibration is divided into two phases 5 4 1 Phase One Display nC the number of calibrations and the respective value are displayed alternately After balancing the scale will move to...

Страница 56: ...needed for servicing the bed Tools Socket set Screwdriver set Engineer s wrench set Hex key bit set Torx bit set Hex key set Knife Set of pliers Torque wrench Hammer nylon hammer Tape measure Crimping...

Страница 57: ...n Date 2019 09 Copyright LINET s r o 2019 Translation LINET s r o 2019 All rights reserved Use this Service Manual in combination with User Manual for Eleganza 4 D9U001GE4 0101 All trademarks and bran...

Отзывы: