D9S001GE4-0101
39/
57
LINIS
iBoard Basic
- To enter LINIS mode simultaneously press (5s)
BEA Inner zone
+
BEA Outer zone
Navigation by (
ZERO
)/(
HOLD
) buttons. Exit by
CLEAR
button.
4.3.1
Description of Linis items displayed on iBoard segment display
LINIS Code
Description
Values
1
AD COL0
AD potentiometer column 0
-----
2
AD COL1
AD potentiometer column 1
-----
4
AD X
AD accelerometer PCB x
-----
5
AD Y
AD accelerometer PCB y
-----
6
AD BR
AD backrest accelerometer
0° - 80° (Chair)
8
AD STOP
AD stop
100-120 (1x STOP), 65-85 (2x STOP)
9
OPER C
Operation codes (see 4.3.3)
XX-YY (XX - function code, YY
– code of periphery)
10
STOP C
Bed stop code (see 4.3.2)
-----
11
SYS ERR
System fault (see 4.1)
-----
12
END-SW
End switches status
xxxxxxx
13
PB SW
Power box SW version
XXYY (9.12->_912; 10.12-> 1012
14
EXP SW
SW version of expansion
-----
15
DU SW
SW version of touchscreen
XXYY (9.12->_912; 10.12-> 1012
End Switches
Function
Position
Value (0 or 1)
Siderails (All)
1
Up=1, Down=0
N/A
2
0
N/A
3
0
Foot column
– lowest
position
4
Activated = 1,
Deactivated = 0
Head column
– lowest
position
5
Activated = 1,
Deactivated = 0
N/A
6
0
N/A
7
0
Содержание Wissner-Bosserhoff ELEGANZA 4
Страница 5: ...D9S001GE4 0101 5 57 Identification of bed sides...
Страница 18: ...D9S001GE4 0101 18 57 Weight module and tensometers S6015189ND...
Страница 31: ...D9S001GE4 0101 31 57 i Drive Power connection scheme...
Страница 34: ...D9S001GE4 0101 34 57 Castors Castors Antistatic castor Castor with lock Central Castor Control...
Страница 35: ...D9S001GE4 0101 35 57 Bed Wiring Diagram...
Страница 45: ...D9S001GE4 0101 45 57 The CAL position displays the calibration constant...
Страница 49: ...D9S001GE4 0101 49 57 5 3 6 4 Positioning the weights for sensors 2 head left sensor...
Страница 51: ...D9S001GE4 0101 51 57 5 3 6 6 Positioning the weights for sensor 3 left foot sensor...
Страница 53: ...D9S001GE4 0101 53 57 5 3 6 8 Positioning the weights for sensor 4 right foot sensor...