D9S001GE4-0101
40/
57
4.3.2
Stop codes
Stop codes
Code
Reason
2
foot control pressed after STOP
3
display button pressed after STOP
4
two buttons pushed simultaneously longer than 1 second
5
keyboard not released after system start
6
main control box temperature exceeded
7
column current overload 0
8
column current overload 1
10
actuator current overload
– backrest
11
actuator current overload
– thighrest
12
stopped due to resistance on the same function
13
short circuiton on analog control (RC detection)
101
GO button not active
102
locked function
111
function blocked during one fault state
113
stopped due to risk of collision when bed is tilted
116
scale overload - > 120 kg
123
absolute scale overload >260 kg