Status (Pop-up window)
Meaning
How to change the status
Optimization process has been stopped
because patient is not properly placed on
the OptiCare X integrated mattress.
Press MUTE Icon to mute acous-
tic alert sounding from SCU.
Position the patient to the centre
of the mattress support platform
according to the indicators on the
foot siderails, head board and
foot board.
Integrated Mattress Fault requiring an
information from the instructions for use
and intervention of a service technician!
See “9.5 Fault Codes (OptiCare X inte-
grated mattress)“ on the page 49.
Press MUTE Icon to mute acous-
tic alert sounding from SCU.
Follow the instructions for the
corresponding fault code display-
ed in the upper right corner of the
pop-up window (
Pxx
) and contact
service department approved
by manufacturer and report the
identifi cation fault code!
Refer to User Manual for Instructions.
Pxx
Mattress optimization fault. Use
siderail indicators to center patient.
P108
Mattress disconnected.
Check air connectors.
P101
RESET
FAULT RESET
Fig. RESET Icon on the pop-up window P101
To reset the displayed fault:
►
Press MUTE Icon on the corresponding pop-up window.
RESET Icon appears on the pop-up window instead of the MUTE Icon.
►
Press RESET Icon.
The fault is reset.
If the fault cause was removed, the fault is eliminated.
If the fault cause was not removed, the fault will return with the corresponding pop-up window.