5.3 Environment Conditions
Use Conditions
Ambient Temperature
10°C - 40°C
Relative Humidity
30% - 75 %
Atmospheric Pressure
795 - 1060 hPa
Storage and Transport Conditions
Ambient Temperature
-40°C - 70°C
Relative Humidity
10% - 100 % (non-condensing)
Atmospheric Pressure
795 - 1060 hPa
5.4 Electromagnetic Compatibility
OptiCare X integrated mattress system is intended for hospitals except for near active HF surgical equipment and the RF
shielded room of a medical system for magnetic resonance imaging, where the intensity of EM disturbances is high.
OptiCare X integrated mattress system has defined no essential performance.
WARNING!
It is recommended to avoid the use of this device next to or in block with other device, because it could lead
to improper operation. If such use is needed, this device and the other equipment should be under surveillan-
ce to verify proper operation. (Does not apply for compatible medical bed from LINET)
List of used cables:
►
Mains cable
, maximum length 6 m
WARNING!
Use of the accessories, converters and other cables, than specified and provided by manufacturer of this Op
-
tiCare X integrated mattress system could lead to increase of electromagnetic emission or lower the electro
-
magnetic immunity of this OptiCare X integrated mattress system and lead to improper operation.
WARNING!
Mobile RF communication device (including end use devices like antenna cables and external antenna)
should not be used closer than 30 cm (12 inches) from any part of this OptiCare X integrated mattress system,
including cables specified by manufacturer. Otherwise this could lead to deterioration of functionality of this
OptiCare X integrated mattress system.
WARNING!
Do not overload the OptiCare X integrated mattress system (SWL), respect the duty cycle (INT.) and consider
chapter 13 Maintenance in order to maintain the basic safety with regard to electromagnetic disturbances for
the expected service life.
5.4.1 Manufacturer Instructions - Electromagnetic Emissions
Emission Test
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Complies