D9U003VSX-0101_00
27
OptiCare X for Multicare X
Fig. Symbol indicating
centre of mattress support
platform
9 Manipulation (OptiCare X for Multicare X)
Preparing OptiCare X for Patient
DANGER!
Risk of suffocation due to air-impermeable mattress cover!
► Use mattress cover correctly.
► The nursing staff are responsible for the safety of the patient on the mattress cover.
WARNING!
Risk of injury when positioning patient on the bed!
Before positioning the patient on the bed:
► Ensure that mattress is completely and correctly inflated.
► Ensure that mattress is correctly secured with safety straps.
► Use of the mattress without the safety straps could impair the mattresses ability to function correctly!
CAUTION!
Material damage due to dampness or contamination!
► Make sure mattress cover has been cleaned and is completely dry (see Cleaning/Disinfection).
Preparation
►
Ensure that mattress is correctly secured with safety straps.
►
Inflate mattress.
►
Put a sheet loosely on the mattress if not prescribed otherwise by qualified personnel.
Fig. Centre of the Multicare X mattress support platform
Position of a Patient on the Bed
►
Put the patient on the mattress. Head of a lying patient points towards head end of the bed.
Position of pelvic area is indicated by symbols on the inner sides of the Multicare X foot siderails,
head board and foot board.