6 Use and Storage Conditions (OptiCare X)
DANGER!
Danger to life due to electric shock!
To ensure the bed’s class I protection against electric shocks:
► Ground the mains.
► Use exclusively Hospital Grade or Hospital Only receptacles for grounding.
Multicare X with OptiCare X integrated mattress system is designed for use in rooms for medical purposes. Electrical installations
must therefore meet local norms laying down the necessary conditions for electrical installations.
►
Disconnect the bed from the mains in exceptional cases (i.e. lightnings, earthquake).
OptiCare X integrated mattress system is not suitable for indoor environments containing flammable gases (except oxy
-
gen cylinders).
6.1 Storage
When SCU is not in use:
►
Log off the mattress.
►
Unplug mains cable.
When mattress is not in use:
►
Unclip all 5 air pipes.
►
Undo webbing strap next to air pipes.
►
Make sure mattress has been cleaned and is completely dry inside and out (see Cleaning/Disinfection).
►
Deflate mattress and leave air connector open (CPR position).
►
Roll mattress up carefully to get air out completely.
Store in a dry and safe place and keep away from sharp objects.
7 Scope of Delivery and Product Variants (OptiCare X)
7.1 Delivery
►
Upon receipt, check that the shipment is complete as specified on the delivery note.
►
Notify the carrier and supplier of any deficiencies or damages immediately as well as in writing or make a note on the
delivery note.
7.2 Scope of Delivery
■
OptiCare X integrated mattress
■
Mattress Cover
□
ELITE
□
ENDURANCE
●
with X-ray Pocket
●
without X-ray Pocket
□
ZONED
●
with X-ray Pocket
●
without X-ray Pocket
■
SCU (System Control Unit)
■
Instructions for use