D9U003VSX-0101_00
35
OptiCare X for Multicare X
25. 12. 25
22:58
MOBILE
PRONE
MAX
OPTIMIZE
Fig. STANDBY SCREEN
STANDBY SCREEN (IN OPTIMIZE MODE)
To perform re-optimization when patient is on the mattress and STANDBY Screen appears in OPTIMIZE Mode:
►
Switch OptiCare X to MOBILE Mode and shortly afterwards back to the OPTIMIZE Mode.
►
If problem is still present and the STANDBY Screen remains displayed, position patient to the middle of the bed according
to the marks on the bed.
►
If problem is still present and the STANDBY Screen remains displayed, patient weight is lower than the minimum limit for
patient (40 kg).