9.2 Controls and Indicators
9.2.1
Control panel - Air2Care 10
The control panel of the SCU serves to control the mattress replacement system and shows errors. Alarms are signaled by illumina-
tion of the indicator
2
.
Fig. SCU Control Panel – Air2Care 10
Position
Control / Indicator
Function
1
Pressure control dial
Adjusting the mattress pressure for more comfort or support of the patient
HI
: High pressure
LO
: Low pressure
2
Low pressure indicator
Indicates low pressure in the mattress (see System errors).
3
Power switch
I
: On
O
: Off
4
Mode selection switch
STATIC:
Fully infl ated mattress. Static mode.
APT:
Alternating pressure therapy. Dynamic mode.
APT – Alternating pressure therapy
Air2Care operates by alternating pressure in a two cell system in a 12 minute interval cycle. This imitates the natural movement of
the patient. During this cycle reduced pressure acts on the patient which helps to prevent and treat pressure sores.
When APT mode is selected:
■
2-cell mattress will infl ate and defl ate in cycles of 12 minutes.
STATIC – Static infl ated mode
Static mode provides a stable surface for the patient when getting into or out of bed or if required when performing nursing proce-
dures. Air cell pressure can be varied using the manual pressure control.
When STATIC mode is selected:
■
SCU will infl ate all cells to the same pressure.