D9U003VA0-0101_04
24
8.1.2 Installation – Mattress Replacement
To secure mattress on the bed:
►
The two separate straps are not used in this situation, and should be removed and stored safely.
►
Ensure that there are no protruding parts or sharp objects on the bed frame to avoid damage to the mattress.
►
Place the deflated Air2Care Replacement Mattress onto the mattress platform, with the air pipe at the left side foot end,
and position so that the mattress is centred. The foot symbol on the mattress cover is located at the foot end of the bed.
►
For each of the 6 base cover straps, pass the webbing (3) around the mattress platform and clip the male part (2) of the
trident buckle into the female part (1). Pull the webbing (3) through the male part (2) to adjust the length of the strap, so
that the mattress is held securely to the frame.
The Replacement mattress can now be connected to the SCU and inflated.
Fig. Fixing of the mattress
8.2 Installation of SCU (System Control Unit)
WARNING!
Risk of injury when putting SCU into service!
►
Make sure that your hands are not trapped between hook and foot board when using spring-loaded SCU hanging
hooks.
►
Make sure that the SCU is installed securely so that it cannot slide or be accidentally knocked off.
WARNING!
Risk of injuring the patient or damaging the accessories due to incorrect putting into service!
►
Ensure the SCU does not collide with any accessories placed on the bed.
CAUTION!
Material damage due to incorrect putting SCU into service!
►
Do not install SCU on linen rack on bed frame.
►
Make sure that the SCU is installed securely so that it cannot slide or be accidentally knocked off.
If foot board of the bed is suitable for hanging SCU:
►
Hold SCU in one hand and unfold hooks on back with the other.
►
Hang SCU on foot board of the bed.
If foot board of the bed is not suitable for hanging SCU:
►
Stand the SCU upright on the floor.
NOTE:
Take extra caution when manipulating the bed or moving around the bed when the SCU is standing on the floor.