IP Protection
Cleaning instrucions
IPX0
Clean with a damp cloth
IPX1
Clean with a damp cloth
IPX2
Clean with a damp cloth
IPX3
Clean with a damp cloth
IPX4
Clean with a damp cloth
IPX5
Wash with a brush and water, but not water under pressure
IPX6
X Wash with a brush and water. The water can be under pressure, but the system must not be
hosed down directly with a high-pressure cleaner. Max. 20oC
IPX6 Washable
Clean by the use of water tunnels
IP Protection degree:
All LINAK products can be cleaned as follow according to their IP protection, which is stated on the
product label.
The IP code specifies the degrees of protection provided by the enclosures. Only the protection
against ingress of water (second characteristic numeral) is specified, ingress of solid foreign objects
(first characteristic numeral) is not specified and therefore replaced by the letter X in the code.
The materials are resistant to the majority of cleaners and disinfectants used in the hospital and
nursing home sector.
Cleaning with a steam cleaner, for example, is not permitted, as it will not be possible to maintain a
minimum distance of 300 mm from the electrical parts.
The following directions regarding cleaners and disinfectants must be complied with:
• They must not be highly alkaline or acidic (pH value 6-8).
• They must not contain caustic agents.
• Their contents must not be able to change the structure of the surface or adhesion of the plastic.
• They must not break down grease.
The following disinfectants/detergents are recommended:
• Neodischer Dekonta, dosage 1-3%, made by CHEMISCHE FABRIK Dr. WEIGERT,
• CCOTRADE RW, dosage 0.5%, made by CCOTRADE GmbH.