Recommendtions:
• Always use locking mechanism and O-ring
• Sockets not used must be fitted with blind plugs to ensure IP Degree
• When using the modular plug with open end, is it up to the customer to maintain the IP degree.
• If the MJB – Gateway is used as open loop, there is a risk of not sending a notification, when no
mains are connected. The MJB – Gateway will only send a notification if mains is missing, when
using a Closed Loop (see functionality description.)
• When the Gateway is used on a system with battery,
the gateway functionality will follow the power
down mode of the control box, see table
• The MJB – Gateway is not to be used as safety, it is only to be used as a guided system.
• It is important to test the specified notification is working in the correct way, before sending the
system to the end user.
• Do not use 2 MJB variants with same device ID on the OpenBus™. This will cause conflicts and
the SDT is not able to identify the different products attached.
• Before the final functional test in the production, it is important that the system is re-powered.
This is to ensure that all items have been detected on the OpenBus™.
• We recommend the end user to make a regular test procedure, in order to prevent failures and
hazardous situations on the system, e.g. squeezed cables. The MJB is not able to detect defective
3
rd
party products.
• LINAK only takes responsibility for LINAK products, not 3
rd
. party products.
Please pay attention to the “Patient Environment” Clause 3.79 - EN60601-1 3
rd
edition: Any volume
in which contact can occur between a patient and parts of the medical equipment or between a
patient and other persons touching parts of the equipment. Justified practice here is 1.5 m. This
volume must be subject to the Risk Analysis.
It is important to inform the customer about this.
• There can be a risk of conflict with other OpenBus™ accessories, such as HB, ACP, etc.
It is therefore recommended to make a system/bit overview.
CB Power mode
Gateway
On mains
Function ok
On battery
Power down
“Wake Up”
No function
Function ok