Example of closed loop:
3
rd
. party site
In this state no notification
Example of Open Loop:
3
rd
. party site
In this state no notification
Features:
• Add-on value for hospital & care bed
• Compact and functional design which allows easy mounting
• 2 connections, Port1 OpenBus™ connection / Port 5 Gateway
• Colour: Light grey RAL 7035
• Protection class IPX6
• Connection to OpenBus™ control box via modular plug cable, 6 wires
• Modular plug to Open end, see precautions.
Options (special article):
• Customised configuration (Configuration number ≥ 100)
• It is possible to define the OpenBus™ signals to be used for activating the relays and / or
change the notification indicator, so that it fits to the customer’s system.
Please contact your Sales Engineer at LINAK A/S for further information.
Usage:
• Compatible with all LINAK OpenBus™ Control Boxes, light & full.
• Contact rating :
Max continuous current:
= 1A
Max switching voltage:
= 48 V (DC)
Max switching capacity:
= 24 VA
Contact resistance:
< 100 mΩ
• Current consumption:
V bus 8V :
< 9 mA
V permanent 40V :
< 14 mA
• Approvals: The MJB has been approved according to IEC60601-1:2005 3
rd
edition, ANSI / AAMI
ES60601-1:2005, 3
rd
edition, and CAN / CSA-22.2 No 60601-1:2008