LIGHTMED TruScan Pro
–
Operator’s Manual Rev. No. 01
Page
23
of
109
3.2
LDU Specifications
3.2.1
Slit Lamp Specifications
Item
Description
Model
CSO SL950 slit lamp
Microscope
Galilean
Pattern scanning
•
Single
•
Line
•
Square (2 x 2, 3 x 3, 4 x 4, 5 x 5)
•
Triangle
•
Circle (single, triple)
•
Equal spacing (2 x 2, 3 x 3, 4 x 4, 5 x 5)
Spot sizes (µm)
•
Single spot: 50, 100, 200, 300, 400 and 500
•
Multi-spot: 100, 200, 300, 400, and 500 selectable on zoom
assembly on slit lamp
Focus plane
All spot sizes parfocal to slit lamp focus plane
Exposure duration (ms)
•
Single spot: 10 to 3,000
•
Multi-spot: 10 to 50
Repeat interval
•
Single spot: selectable from OFF, 10 to 3,000 ms and greater-
than or equal to (≥) exposure duration, in the same discrete
steps as exposure duration
•
Multi-spot: OFF
Slit lamp
The TruScan module is mounted between the binocular and
objective lens
Beam divergence
Cone angle of 20°
Magnification set
6-step drum rotation
Eyepiece
12.5X
Magnification ratio
•
5X
•
8X
•
14X
•
24X
•
36X
PD range (mm)
48.5 to 80
Diopter adjustment
±8
Slit illumination
LED XLamp® XM-L2 2.85V 10W
Slit width (mm)
0 to 14 (SL950)
Slit length (mm)
1 to 14
Slit apertures (mm)
•
0.3
•
5.5
•
9
•
14
Slit angles
±90°
Filters
•
Blue
•
Red
•
Green (red-free)