LIGHTMED TruScan Pro
–
Operator’s Manual Rev. No. 01
Page
105
of
109
Section 8
User Maintenance
The TruScan Pro laser system is designed to require minimal maintenance. There are several
simple routine procedures that must be carried out by the user, but aside from that, there are no
user maintenance requirements.
The Manufacturer, however, does recommend that the TruScan Pro laser systems be serviced
by an authorized Service Engineer every 12 months. During this preventive maintenance (PM)
service visit, the laser system will be calibrated and aligned while verifying general operating
functions.
It is a requirement that on an annual basis (every 12 months), the TruScan Pro laser system
be calibrated and have its earth leakage current and earth resistance measured according
to EN60601-1. These procedures can only be performed by an authorized Service Engineer
approved by the Manufacturer.
If at any time the user has any concerns about any aspect of the laser system, they are urged to
contact the authorized representative or the Manufacturer in order to decide on a suitable course
of action.
Refer to Section 9 of this Manual for specific national requirements for EU countries.
The following procedures are recommended by the Manufacturer for the user to follow:
1.
Clean the external surfaces of the laser using a damp, but not dripping, cloth. Use a mild
cleaning agent and do not use any solvents. Do not spray or pour any cleaning agents directly
onto the laser system. Avoid touching the optics, as there is a specific procedure to clean
them (Step 2 below). Use a dry cloth afterwards or allow it to air dry. This procedure should
be carried out often, at least once every three months. When the laser system is not in use,
keep it covered using the dust cover provided by the Manufacturer.
2.
Clean the optics (objective lens, illumination tower prism, the eyepieces, and other slit lamp
accessory) by gently wiping across the optics using the moistened end of a cotton tip (Q-tip)
or a folded lens tissue (Kodak or similar) with 100% methanol or ethanol. Use only one Q-tip
of lens tissue per wipe, then discard it. Repeat with a new Q-tip or lens tissue until the surfaces
of the optics are clean. Never wipe an optic with a dry Q-tip or lens tissue as this will scratch
the optics.