Lighthouse REMOTE Liquid Particle Counter (4-20mA output) Operating Manual
4-12
248083296-1 Rev 3
<rb7500>Range2= 7500.
SET SAMPLE TIME (seconds)
Note:
Sample times <5
seconds are not
recommended.
<ta#>
Sets sample time in # seconds, where # >1 and nnnn < 3600.
Type:
<ta30>
Response:
<ta30>Sample Time = 30 secs
SET ALARM CHANNEL
<aa#>
Sets the alarm channel to channel 1 or 2.
Type:
<aa1>
Response:
<aa1>Alarm Channel= 1
SET ALARM THRESHOLD
<ab#>
Sets Alarm High count threshold for the configured alarm channel.
Value # must be less than range in order to set the threshold. Set the
value to 0 (zero) to disable alarm.
If the count for the configured alarm channel exceeds the threshold
then:
•
The red SERVICE indicator on the instrument blinks.
•
The EXTERNAL ALARM relay, described below, is set.
•
The alarm alert will be seen on the ASCII terminal as illustrated
next:
Содержание RLPC 0.1mm
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counters 4 20mA Output Operating Manual...
Страница 8: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual t iv 248083296 1 Rev 3...
Страница 10: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual ii 248083296 1 Rev 3...
Страница 20: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 2 8 248083296 1 Rev 3...
Страница 29: ...Get Started 248083296 1 Rev 3 3 9 3 Wire System Figure 3 7 Three Wire Communication Diagram...
Страница 30: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 3 10 248083296 1 Rev 3...
Страница 50: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 5 4 248083296 1 Rev 3...
Страница 54: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual B 2 248083296 1 Rev 3...
Страница 59: ...00...
Страница 60: ......